Objective Assessment of Postoperative Results of Intraocular Lenses
Authors:
J. Nekolová; H. Langrová; P. Rozsíval; N. Jirásková
Authors place of work:
Oční klinika Fakultní nemocnice Hradec Králové
přednostka prof. MUDr. Naďa Jirásková, Ph. D., FEBO
Published in the journal:
Čes. a slov. Oftal., 72, 2016, No. 3, p. 66-70
Category:
Původní práce
Summary
Purpose:
To prospectively compare postoperative results of two premium intraocular lenses EnVISTA (Bausch and Lomb) and AcrySof IQ (Alcon), focussing on glistenings and posterior capsule opacification. The evaluation of glistenings was done using Image J software and posterior capsule opacifications were quantified with OSCA system.
Methods:
Twenty patients (7 men and 13 women) with bilateral cataract were included. EnVista intraocular lens (IOL) was implanted in one eye and AcrySof IQ IOL in the second eye of each patient. Objective evaluation methods were used for assessment. Glistenings was quantified with ImageJ software and PCO using the Open-Access Systematic Capsule Assessment (OSCA) system (Devised by Aslam TM, Edinburgh, United Kingdom). Complete ophthtalmological evaluation including evaluation after pupil dilatation was done and digital images of intraocular lenses were obtained. The results of 2-, 4-, 6- and 12-month follow-up were compared.
Results:
Twenty patients were analyzed 2 months, 16 patients 4 months, 14 patients 6 months and 13 patients 12 months after cataract surgery. There was only minimal difference in best corrected visual acuity between EnVista and AcrySof group. The glistenings in the EnVista IOLs was objectively lower than in the AcrySof IOLs during whole follow-up period. In contrast to PCO, in eyes with AcrySof IOL was lower PCO score.
Conclusion:
Development of new materials and techniques of cataract surgery is the topic of ophthtalmologists worldwide. Reduction of glistenings and PCO is one of the main aims, objective measurements is important part of assessment of postoperative results after cataract surgery.
Key words:
glistenings, posterior capsule opacification, EnVista, AcrySof IQ, Image J software, OSCA system
Zdroje
1. Adámková H. Novák J.: Sekundární katarakta, její prevence a léčba. Část první: Prevence vzniku sekundární katarakty. Čes a slov Oftal, 62; 2006: 230–236.
2. Aslam, TM., Patton, N., Rose, ChJ.: OSCA: a comprehensive open-access system of analysis of posterior capsular opacification, BMC Oftalmol, 6; 2006: 30.
3. Beheregaray S., Yamamoto, T., Hiraoka T. et al.: Influence on visual function of forward light scattering associated with subsurface nanoglistenings in intraocular lenses. J Cataract Refract Surg, 40; 2014: 1147–1154.
4. Bessen-Miyajima H., Minami K., Yoshino M., et al.: Sufrace light scattering and visual function of diffractive multifocal hydrophobic acrylic intraocular lenses 6 years after implantation. J Cataract Refract Surg, 39; 2013: 1729–1733.
5. Colin J., Orignac I.: Glistenings on Intraocular Lenses in healthy Eyes: Effects and Associations. J Refract Surgery, 36; 2010: 1398–1420.
6. Congdon, N.: Three-year study compares acrylic and silicone IOLs. Ophthalmology Times. 1997, 15, 14-15. a Mandle, M. Acrylic lenses cause less posterior capsular opacification than PMMA or silicone. Ocular Surgery News, 8; 1996: 9.
7. Deepinder K.D., Mamalis N., Olson R.J., et al.: Visual significance of glistenings seen in the AcrySof intraocular lens. J Cataract Refract Surg, 22; 1996: 452–457.
8. Gunenc U., Oner H., Tongal S.: Effects on visual function of glistenings and folding marks in AcrySof intraocular lenses. J Cataract Refract Surg, 27; 2001: 1611–1614.
9. Hayashi K., Hirata A., Yoshida M., et al.: Long-Term Effect of Surface Light Scattering and Glistenings of Intraocular Lenses on Visual Function. Am J Ophthalmol, 154; 2012: 240–251.
10. Chang A., Behndig A., Ronbeck M., et al.: Comparison of posterior capsule opacification and glistenings with 2 hydrophobic acrylic intraocular lenses: 5- to 7-year follow-up. J Cataract Refract Surg, 39; 2013: 694–698.
11. Jirásková, N., Rozsíval, P.: Metody hodnocení zkalení zadního pouzdra po operaci katarakty. Čes. a slov. Oftalmol, 60; 2006: 155–157.
12. Linnola, R.J.: Sandwich theory: Bioactivity-based explanation for posterior capsule opacification. J. Cataract. Refract. Surg, 23; 1997: 1539–1542.
13. Miyata A., Yaguchi S.: Equilibrium water content and glistenings in acrylic intraocular lenses. J Cataract Refract Surg, 30; 2004: 1768–1772.
14. Nekolová, J., Pozlerová, J., Jirásková, N. et al.: Pooperační výsledky expandibilní nitrooční čočky ACQUA (Mediphacos). Čes a Slov Oftalmol, 64; 2008: 87–90
15. Nekolová, J., Jirásková, N., Pozlerová, J., Rozsíval, P.: Three-year follow-up of posterior capsule opacification after AquaLase and NeoSoniX phacoemulsification. Am J Ophthalmol, 148; 2009: 390–395.
16. Nekolová, J., Pozlerová J., Jirásková N., Rozsíval P.: Opacity zadního pouzdra u pacientů s diabetes mellitus 2. typu. Čes. a slov. Oftal, 64; 2008: 193–196
17. Pagnoulle Ch., Bozukova D., Gobin L., et al.: Assessment of new-generation glistening-free hydrophobic acrylic intraocular lens material. J Cataract Refract Surg, 38; 2012: 1271–1277.
18. Spalton, D.J., Maxwell, W.A.: Acrylic lenses minimize PCO progression: British study. Ocular Sugery News, 14; 1996: 34.
19. Stepanov, S., Nekolová, J., Jirásková, N., Rozsíval, P.: Long-term follow-up of posterior capsule opacification after AquaLase and NeoSoniX phacoemulsification, Biomed Pap Med Fac Univ Palacky Olomouc Czech repub, v tisku.
20. Tognetto D et al.: Glistenings in foldable intraocular lenses. J Cataract Refract Surg, 18; 2002: 1211–1216.
21. Ursell, P.G., Spalton, D.J., Pande, M.V., Hollick, E.J., Barman, S., Boyce, J., Tilling, K.: Relationship between intraocular lens biomaterials and posterior capsule opacification. J. Cataract Refract. Surg, 24; 1998: 352–360.
22. Werner, L.: Glistenings and surface light scattering in intraocular lenses. J Cataract Refract Surg, 36; 2010:1398-1420.
Štítky
OftalmológiaČlánok vyšiel v časopise
Česká a slovenská oftalmologie
2016 Číslo 3
- Myasthenia gravis: kombinace chirurgie a farmakoterapie jako nejefektivnější modalita?
- Cyklosporin A v léčbě suchého oka − systematický přehled a metaanalýza
- Pomocné látky v roztoku latanoprostu bez konzervačních látek vyvolávají zánětlivou odpověď a cytotoxicitu u imortalizovaných lidských HCE-2 epitelových buněk rohovky
- Konzervační látka polyquaternium-1 zvyšuje cytotoxicitu a zánět spojený s NF-kappaB u epitelových buněk lidské rohovky
- Dlouhodobé výsledky lokální léčby cyklosporinem A u těžkého syndromu suchého oka s 10letou dobou sledování
Najčítanejšie v tomto čísle
- Výsledky implantace trifokální torické čočky u pacientů s kataraktou
- Místní otočné laloky v okuloplastice
- Prvé skúsenosti s korekciou presbyopie femtosekundovým laserom metódou INTRACOR
- Objektivizace pooperačních výsledků nitroočních čoček