A performance tester of defibrillator accumulators for clinical purposes


Defibrillator battery life length depends on the way of the utilization of the defibrillator, the battery type and effectiveness of the defibrillator’s internal charging mechanism, as well as access of the personal. This paper describes the basics of energy discharging of the accumulator during the defibrillator clinical utilization. The main parameters and its monitoring are described as well. The methods of testing have been designed. The main part of this thesis describes the construction of the device for the testing of defibrillator’s accumulators and its use in the Department of medical devices in General faculty hospital in Prague. Several batteries were tested, the old an often used as well as the new batteries bought from the producer of the defibrillators. Based on the principles of lead-acid battery there is a predicted improvement in the acu-pack capacity while periodically being tested using our testing routine.

Keywords:
defibrillator, lead-acid battery, measurement of capacity, performance tester


Autoři: Jan Suchomel;  Ondřej Čadek
Působiště autorů: Czech Technical University in Prague, Faculty of Biomedical Engineering, Kladno, CZ
Vyšlo v časopise: Lékař a technika - Clinician and Technology No. 4, 2012, 42, 28-31

Souhrn

Defibrillator battery life length depends on the way of the utilization of the defibrillator, the battery type and effectiveness of the defibrillator’s internal charging mechanism, as well as access of the personal. This paper describes the basics of energy discharging of the accumulator during the defibrillator clinical utilization. The main parameters and its monitoring are described as well. The methods of testing have been designed. The main part of this thesis describes the construction of the device for the testing of defibrillator’s accumulators and its use in the Department of medical devices in General faculty hospital in Prague. Several batteries were tested, the old an often used as well as the new batteries bought from the producer of the defibrillators. Based on the principles of lead-acid battery there is a predicted improvement in the acu-pack capacity while periodically being tested using our testing routine.

Keywords:
defibrillator, lead-acid battery, measurement of capacity, performance tester


Zdroje

[1] Cenek, M., et al. Akumulátory: Od principu k praxi. FCC Public. Praha. 2003. ISBN 80-86534-03-0.

[2] Kaválek, J., 555C: Příručka pro konstruktéry. Epsilon. Paha. 1994. ISBN 90-900035-5-9.

[3] ZOLL M-Series configuration guide. ZOLL Medical Corporation. 2007.

[4] Precision Low Power Instrumentation Amplifier INA 128. Datasheet. Texas Instruments. 2011.