Home measurement of blood pressure: present problems and perspective improvements
The most frequently performed health test is the measurement of blood pressure. Popularity of home measurement increased substantially with the introduction of inexpensive automatic instruments to the market. Accurate measurement of blood pressure is important for diagnosis and management of hypertension. Inaccuracies of measurement with automatic oscillometric instruments are caused by wrong size cuff and by errors in algorithmic measurement. The authors developed several perspective improvements in the measurement of systolic pressure. An experimental system for acquisition and processing of arterial pressure pulses facilitated the development of these methods. More accurate measurement of the systolic pressure was achieved with a dual-cuff method. A new method of wrist-cuff pulse analysis facilitated estimation of hemodynamics. Estimation of hemodynamics simultaneously with measurement of blood pressure provides the physician with more complete picture of the type of hypertension and it facilitates better diagnosis and management. An inexpensive commercial instrument based on the methods developed by the authors could be used by the patient in home care.
Keywords:
blood pressure, home measurement, oscillometric method, empiric algorithms, experimental system, hemodynamics of hypertension
Autoři:
Jiri Jilek 1; Milan Stork 2
Působiště autorů:
Carditech, Culver City, California, USA
1; Faculty of Electrical Engineering/RICE, University of West Bohemia, Plzen, CZ
2
Vyšlo v časopise:
Lékař a technika - Clinician and Technology No. 4, 2012, 42, 36-40
Souhrn
The most frequently performed health test is the measurement of blood pressure. Popularity of home measurement increased substantially with the introduction of inexpensive automatic instruments to the market. Accurate measurement of blood pressure is important for diagnosis and management of hypertension. Inaccuracies of measurement with automatic oscillometric instruments are caused by wrong size cuff and by errors in algorithmic measurement. The authors developed several perspective improvements in the measurement of systolic pressure. An experimental system for acquisition and processing of arterial pressure pulses facilitated the development of these methods. More accurate measurement of the systolic pressure was achieved with a dual-cuff method. A new method of wrist-cuff pulse analysis facilitated estimation of hemodynamics. Estimation of hemodynamics simultaneously with measurement of blood pressure provides the physician with more complete picture of the type of hypertension and it facilitates better diagnosis and management. An inexpensive commercial instrument based on the methods developed by the authors could be used by the patient in home care.
Keywords:
blood pressure, home measurement, oscillometric method, empiric algorithms, experimental system, hemodynamics of hypertension
Zdroje
[1] Spinar J, Vítovec J, Zícha J a kolektiv: Hypertenze: diagnostika a léčba. Grada Publishing 1999.
[2] Jones D, Appel JL, Sheps SG, Rocella EJ, Lenfant C: Measuring Blood Pressure Accurately. JAMA 2003;289:1027-1030.
[3] O’Brien E, Petrie , Littler W: An outline of the revised British Hypertension Society protocol for the evaluation of blood pressure monitoring devices. Blood Press Monit. 2002;7:17-30.
[4] O’Brien E, Pickering TG, Asmar R, Myers MG, Staessen J: Working Group on Blood Pressure Monitoring of The European Society of Hypertension: International Protocol for validation of blood pressure measuring devices in addults. Blood Press Monit. 2002;7:3-17.
[5] Gerin W, Schwartz AR, Pickering TG, Davidson KW, Bress J, O’Brien E: Limitations of current validation protocols for home blood pressure monitors for individual patients. Blood Press Monit. 2002;7:313-318.
[6] Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves J: Recommendations for Blood Pressure Measurement in Humans and Experimental Animals. Hypertension 2005;45:142-161.
[7] Marks LA, Groch A: Optimizing cuff width for noninvasive measurement of blood pressure. Blood Press Monit. 2000;5:153-156.
[8] O’Brien E, Littler Wade Swiet M, Padfield PD, Dillon MJ: Blood pressure measurement: Recommendations of the British Hypertension Society. BMJ Books, 1997.
[9] Jilek J, Stork M: Wrist Cuff Width Alters Amplitude Envelope of Oscillometric Blood Pressure Waveforms. Proceedings of Applied Electronics 2008 International Conference, September 10-11, Plzen, CR, University of West Bohemia.
[10] Kikuya M, Chonan K, Imay Y, Goto E, Ishii M: Accuracy and reliability of wrist-cuff devices for self-measurement of blood pressure. J Hypertens 2002;20:573-578.
[11] Jilek J, Fukushima T: Oscillometric Blood Pressure Measurement: The Methodology, Some Observations, and Suggestions. Biomed Instrum Technol 2005;39:237-241.
[12] Geddes LA. Characterization of the oscillometric method for measuring indirect blood pressure. Ann Biomed Eng Comput 1982;10:271-280.
[13] van Popele NM, Bos WJ, de Beer N, van der Kulp D, Hofman A, Grobbee DE, Wilteman J. Arterial stiffness as underlying mechanism of disagreement between an oscillometric blood pressure monitor and a sphygmomanomeeter. Hypertension 2000;36:484-488.
[14] Czarkowski M, Mikulska M, Zebrowski M, Baran A, Rozanowski K. Venous blood redistribution alters the accuracy of Omron HEM-705CP. Blood Press Monit 2004;9:9-12.
[15] Borow KM, Newburger JW. Noninvasive estimation of central aortic pressure using the oscillometric method for analyzing systemic artery pulsatile blood flow. Comparative study of indirect systolic, diastolic and mean brachial artery pressure with simultaneous direct ascending aortic pressure measurement. Am Heart J. 1982;103:879.
[16] Jones DW, Frohlich ED, Grim CM, Grim CE, Taubert KA. Mercury sphygmomanometers should not be abandoned: an advisory statement from the Council for High Blood Pressure Research, American Heart Association. Hypertension 2001;37:185-186.
[17] Jilek J, Stork M. Systém pro neinvazivní vyšetření krevního tlaku a hemodynamiky. Lekař a Technika 2004; 35:33-38.
[18] [Jilek J, Stork M. An Experimeental System for Estimation of Blood Pressures and Hemodynamics from Oscillomeetric Waveforms. Proceedings of Applied Electronics 2003 International Conference, September 10-11, Plzen, CR, University of West Bohemia.
[19] Staessen JA. Blood pressure measuring devices: time to open Pandora’s box and regulate. Hypertension 2000; 35:1037.
[20] Laragh JH, Brenner BM. Hypertension: Pathophysiology, Diagnosis and Management, Raven Press, 1995.
[21] Taler SJ, Textor SC, Augustine JE. Resistant Hypertension: Comparing Hemodynamic Management to Specialist Care. Hypertension 2002; 39:982-988.
[22] Scardo JA, Vermillion ST, Hogg BB, Newman RB. Hemodynamic effect of oral nifedipine in preeclemptic hypertensive emergencies. Am J ObGyn 1996; 175:336-340.
Štítky
BiomedicínaČlánok vyšiel v časopise
Lékař a technika
2012 Číslo 4
Najčítanejšie v tomto čísle
- Úskalí termografického měření v medicíně
- Fantom pro diagnostický ultrazvuk a dopplerovské vyšetření
- Vliv ultrazvuku na účinnost fotodynamické terapie – in vitro studie
- Vliv poruch sondy sonografu na kvalitativní parametry ultrazvukového B-obrazu