Cannula and circuit management in peripheral extracorporeal membrane oxygenation: An international survey of 45 countries
Authors:
Taressa Bull aff001; Amanda Corley aff001; India Lye aff001; Amy J. Spooner aff001; John F. Fraser aff001
Authors place of work:
Critical Care Research Group, The Prince Charles Hospital and University of Queensland, Chermside, Queensland, Australia
aff001; Adult Intensive Care Services, The Prince Charles Hospital, Chermside, Queensland, Australia
aff002
Published in the journal:
PLoS ONE 14(12)
Category:
Research Article
doi:
https://doi.org/10.1371/journal.pone.0227248
Summary
Effective and safe practices during extracorporeal membrane oxygenation (ECMO) including infection precautions and securement of lines (cannulas and circuits) are critical to prevent life-threatening patient complications, yet little is known about the practices of bedside clinicians and data to support best practice is lacking. Therefore, the aim of this study was to identify and describe common line-related practices for patients supported by peripheral ECMO worldwide and to highlight any gaps for further investigation. An electronic survey was conducted to examine common line practices for patients managed on peripheral ECMO. Responses were obtained from 45 countries with the majority from the United States (n = 181) and United Kingdom (n = 32). Standardised infection precautions including hand hygiene, maximal barrier precautions and skin antisepsis were commonplace for cannulation. The most common antisepsis strategies included alcohol-based chlorhexidine gluconate (CHG) for cannula insertion (53%) and maintenance (54%), isopropyl alcohol on circuit access ports (39%), and CHG-impregnated dressings to cover insertion sites (36%). Adverse patient events due to line malposition or dislodgement were reported by 34% of respondents with most attributable to ineffective securement. Centres ‘always’ suturing peripheral cannula sites were more likely to experience a cannula adverse event than centres that ‘never’ sutured (35% [95% CI 30, 41] vs 0% [95% CI 0, 28]; Chi-square 4.40; p = 0.04) but this did not meet the a priori significance level of <0.01. An evidence-based guideline would be beneficial to improve ECMO line management according to 78% of respondents. Evidence gaps were identified for antiseptic agents, dressing products and regimens, securement methods, and needleless valves. Future research addressing these areas may provide opportunities for consensus guideline development and practice improvement.
Keywords:
Pediatrics – Blood – Respiratory infections – Surveys – Adverse events – Alcohols – Hygiene – Drug administration
Introduction
Extracorporeal membrane oxygenation (ECMO) is an invasive method of cardiopulmonary support increasingly used in intensive care units worldwide for patients with respiratory and/or cardiac failure[1] and is high-risk for complications. The use of large-bore cannulas and associated circuit tubing (‘ECMO lines’) permit vascular access and blood flow through the extracorporeal system. Peripheral cannulations are commonly employed as they allow rapid bedside access to ECMO support using percutaneous methods[2]. The success of ECMO is dependent on effective placement and securement of these lines, among other factors. Ineffective securement can contribute to blood flow impairments, compromising patient haemodynamics and oxygenation, or cannula dislodgement which can result in air embolism or catastrophic bleeding[3]. At worst, fatal outcomes occur[4, 5].
Nosocomial infections, which include ECMO cannula-related infections (CRI), are among the most frequent complications in ECMO and cause patient morbidity[6–10] and mortality[11–13], as well as mechanical circuit dysfunction[14] which adds to patient complications. CRIs are often under-recognised contributors to infectious complications during ECMO[15]. In the limited number of studies reporting ECMO CRIs specifically, these infections account for 10% to 18% of infectious episodes[1, 16–18] and have an incidence of up to 17.2 per 1000 ECMO days[1], suggesting that they are of significant concern. Moreover, if the patient has a primary bloodstream infection (BSI) the ECMO cannula should be implicated as the culprit if the causative pathogen cannot be located in another organ, according to Messika et al.[15]. Unlike other invasive lines that undergo routine replacement for suspected or proven infections, ECMO CRIs pose a significant clinical dilemma due to the difficult and hazardous nature of line replacement[19, 20].
Critically, prevention of many of the possible complications in ECMO relies on effective practices to maintain well secured, infection-free lines, yet data to guide clinicians is lacking. Data limited to the United States shows variability in cannula insertion and maintenance practices among ECMO centres[21]. Available recommendations[20, 22] are based largely on evidence only partially applicable to ECMO lines and there is insufficient high-quality evidence specifically to guide effective line securement, dressing and access practices as a means to prevent dislodgement and/or infection in ECMO.
A better understanding of contemporary bedside practices may contribute towards the development of quality data to guide management of ECMO patients. Our aim was to identify and describe common line-related practices for patients supported by peripheral ECMO worldwide and to highlight any gaps for further investigation.
Materials and methods
A cross-sectional, descriptive survey was conducted over a one-month period using an online survey platform (SurveyMonkey Inc., San Mateo, C.A., USA). Approval to conduct the study as an audit exempt from full ethical review was obtained by the relevant hospital human research ethics committee (HREC/15/QPCH/49). The study was endorsed by the International ECMO Network (ECMONet). Directors and coordinators from 396 ECMO centres in 51 countries that were members of the Extracorporeal Life Support Organization (ELSO) were invited by email to participate using an electronic survey link. These positions were specifically targeted as they would most likely be aware of local policy and practice. The invitation included instructions requesting only one director or coordinator complete the survey from their respective centre and to select the responses which best described that centre’s standard practices. A reminder email was sent two weeks after survey activation then resent one week before the survey closed. Participation was voluntary and anonymous, and submission verified consent to participate.
Data were collected through a 30-item investigator-developed questionnaire covering questions on centre characteristics and cannulation methods, peripheral lines (cannula insertion, dressings and securement) and circuit access, and clinical practice guidelines (refer to S1 Appendix). The questionnaire tool was assessed by an expert panel of multidisciplinary ECMO clinicians from our centre including doctors, nurses and perfusionists for feasibility, readability, understandability, ease of response, and content validity. The tool was pilot tested at two different time points by 10 ECMO clinicians to examine test-retest reliability and internal consistency. Inter-rater agreement was good with an average proportion of agreement of 0.83 across questions and average Gwet's agreement coefficient[23] of 0.73. Inter-rater agreement analyses were performed using the ‘KAPPAETC’ module[24] in the Stata Statistical Software package (StataCorp LP, version 13, College Station, Texas, USA).
Survey data were collected anonymously and downloaded into a statistical software package (SPSS Statistics for Windows, version 24, IBM Corp, Armonk, N.Y., USA) for analysis using descriptive and inferential statistics. Frequencies and percentages were calculated for the responses to each question. To test the null hypothesis of no difference in proportions for selected variables, a Chi-square or Fisher’s Exact test was used dependent on the number of cases in the tested categories. To account for multiple significance testing and the exploratory nature of this analysis, Bonferroni adjustments were made to reduce the significance level to <0.01. Free-text responses to the question relating to cannula malposition and dislodgement were manually categorised and coded into the reason leading to the adverse patient outcome.
Results
Centre characteristics and peripheral cannulation methods
A total of 391 individual responses were received among 45 countries from 911 distributed surveys; two respondents did not provide any clinical data and were excluded from analysis, yielding an overall response rate of 43%. The possibility of double enrolments in centres cannot be ruled out. Approximately half (56%) of the responses came from the Americas and one-quarter from Europe (28%); the greatest number received by country were from the United States (n = 181) and the United Kingdom (n = 32). Eighty-five percent of ECMO centres were in an academic teaching hospital and 43% had been providing ECMO for more than 15 years. The percentage of responses by world region and centre characteristics is shown in Table 1. Respondent centres included neonatal, paediatric and adult patient populations receiving ECMO for respiratory, cardiac and cardiopulmonary resuscitation reasons (Table 2).
The most common peripheral cannulation method was percutaneous Seldinger technique for venovenous access and return (60% each) and venoarterial access (41%); and direct cut-down method for venoarterial return (43%). The most commonly reported cannulation vessels were jugular vein with double-lumen cannula for neonatal and paediatric venovenous ECMO (85% and 65% respectively); femoral-jugular vein (39%) for adult venovenous ECMO; jugular vein-carotid artery for neonate (96%) and paediatric (64%) venoarterial ECMO; and femoral vein-femoral artery (81%) for adult venoarterial ECMO.
Infection prevention for peripheral cannulation, cannula dressing and circuit access
Hand hygiene was routinely performed by 89% of respondents, and 91% prepped the skin with antiseptic prior to peripheral cannula insertion. During peripheral cannulation, respondents indicated the use of maximal barrier precautions as follows: 90% used cap/hat, mask and sterile gown; 91% used sterile gloves; and 85% used a sterile full body drape. Three-quarters of respondents stated that a sterile occlusive dressing was used to cover the insertion site.
Hand hygiene was performed by 83% of respondent centres when replacing dressings. Clinical practice during cannula dressing changes was identified as: use of a cap/hat in almost half of respondents (47%), over two-thirds used a mask (68%); and just over three-quarters used sterile gloves (79%), while 13% used non-sterile gloves. Forty percent stated that an aseptic ‘no-touch’ technique was used during dressing replacement. Prior to accessing circuit ports, reported precautions undertaken included hand hygiene (77%), non-sterile glove (60%) and sterile glove (20%) use.
The preferred antiseptic was chlorhexidine gluconate (CHG) in alcohol for skin antisepsis prior to cannula insertion and during dressing changes (53% and 54% respectively). Isopropyl alcohol swab was most common for disinfection of circuit access ports (39%). Table 3 shows the top five preferred antiseptics used for cannula insertion and maintenance, and circuit port access.
Peripheral cannula infection surveillance
Most centres monitored for cannula site infection by either routine visual inspection through the dressing or during dressing change (84%), swab or culture on signs of infection (31%) or palpation through the dressing (18%). Routine collection of site swabs or subcutaneous needle aspiration of suspicious cannula sites was infrequently performed.
Peripheral cannula dressing and line securement
The frequency of routine dressing replacement was most commonly ‘as required’ (i.e. soiled or bloody) or every 1–3 days (both 29%), followed by daily (20%). No centres reported that dressings were left in situ for greater than seven days. Three-quarters of peripheral cannula sites were dressed with a transparent semi-permeable dressing–40% of these with a plain transparent dressing; 21% with an integrated (1-piece) CHG-impregnated dressing; and 15% with a CHG disk plus transparent dressing over the top. Gauze and tape dressing was reported by 16% of respondents. There was widespread uptake of CHG-containing products for cannula dressing with 28 countries reporting their use; the highest users of such products being the Americas (40% of peripheral cannula dressings contained CHG) compared with the lowest being Australia/New Zealand (21%).
Peripheral ECMO cannulas were almost always sutured at the insertion site with 93% of respondents reporting this practice. Comparing regional variations in suturing practice, the Americas (96%; 95% CI 92, 98) and Europe (96%; 95% CI 90, 99) reported the most frequent use of this securement method compared with Australia/New Zealand reporting the lowest (68%; 95% CI 46, 85; Chi-square 50.6; p <0.0001). When examining the securement method primarily used along the length of the ECMO line/circuit tubing to prevent decannulation or dislodgement, suturing directly to the skin (i.e. leg, neck) was the most prevalent method (45%) followed by securing to a bed or other fixed object (including use of taping, clipping or tubing holder device, 19%). Use of a sutureless securement device (SSD) or adhesive fabric tape/bandage was 16% and 10% respectively. Forty-eight percent routinely used two fixation points along femoral line while approximately one-quarter used more than two (26%) or one (21%) point of fixation. Table 4 describes regional variations in methods of peripheral ECMO line securement.
Concerningly, one-third (34%) of respondents reported the occurrence of a cannula/line malposition, dislodgement or accidental decannulation leading to an adverse patient outcome at their centre in the last five years. Over half of the respondents to this question provided narrative data (61 respondents described 71 adverse events) from which the leading cause of malposition, dislodgement or decannulation was inadequate securement (Table 5).
Associations between cannula/line adverse events (malposition or dislodgement/decannulation) and bedside practices revealed that adult centres handling greater than 30 cases per year reported more adverse events than those performing 1–6 cases per year (56% [95% CI 46, 66] vs 22% [95% CI 13, 35]; Chi-square 16.17; p <0.001). There were no significant differences in ECMO volume per year and a cannula adverse event in neonates and paediatric centres. Counterintuitively, 34% (95% CI 30, 41) of centres which always sutured at the cannula insertion site reported adverse line events whereas, in contrast, centres which never sutured reported none (95% CI 0, 28), however this failed to meet the a priori statistical significance level (Chi-square 4.40; p = 0.04). Length of ECMO experience had no association with adverse line events with centres providing ECMO for 0–2 years reporting comparable cannula adverse event rates than centres with over 15 years of ECMO experience (17% [95% CI 5, 40] vs 36% [95% CI 29, 44; Fishers Exact test, 0.12; p = 0.116). Similarly, there were no significant associations between a cannula adverse event occurring and whether the ECMO centre was in an academic teaching hospital or not (Chi-square 2.61; p = 0.272).
Circuit access
The number of times for daily circuit access was 0–5 for more than half (54%) the respondent centres. Access occurred to perform blood sampling for oxygenator blood gas analysis (93%) and three-quarters of respondents reported (76%) undertaking haemodialysis or haemofiltration via the circuit. Table 6 describes circuit access practices.
Circuit connection methods were reported most frequently as luer-lock for syringe (55%), infusion line (38%), and extracorporeal line (for haemodialysis/haemofiltration or plasmapheresis; 65%), whereas connections using a needleless valve were less frequent (syringe, 27%; infusion line, 18%; extracorporeal line, 8%). Up to 13% of respondents reported either connection method was used.
Guidelines for ECMO line management practices
Over three-quarters (76%) of the respondents had a written policy or guideline for bedside ECMO line management while the remainder did not (22%) or were unsure (3%) if a policy existed at their centre. Most (78%) believed that an international evidence-based guideline for ECMO line management would be beneficial to inform or improve bedside practices.
Discussion
Results from this survey provide insights into bedside management of peripheral ECMO lines among 45 countries with evidence of practice variation in infection precautions, dressings, securement and connections. A number of management aspects appear inconsistent with available recommendations including those from the ELSO Infectious Disease Task Force[22].
Infection precautions and surveillance
The use of standardised infection prevention practices relevant to central venous catheters[25] appears to be commonplace during ECMO cannula insertion with the vast majority of centres reporting adherence to maximal barrier precautions, performing hand hygiene, disinfecting the patient’s skin with alcohol-based antiseptic, and applying a sterile full-body drape. Although hand hygiene had a high reported compliance for cannula insertion and dressing replacement, compliance was lower prior to circuit access with nearly one-in-four respondents reporting that hand hygiene was not routinely performed. This may reflect perceptions or attitudes that the cannulas represent a greater risk for microbial contamination than does entering the circuit ports. Should this be the case, it contravenes recommendations of the ELSO Infectious Disease Task Force[22] to treat ECMO circuits as protected central lines as both cannula handling and port accessing involve breaking the sterile circuit and potentially allow microbial contamination. However, behavioural reasons may influence hand hygiene practices such that compliance may be poorer depending on the clinical context (e.g. emergency situation), which we did not specifically survey.
Similarly, avoiding unnecessary circuit access is a recommended strategy to reduce the risk for infection[22]. Our results demonstrate that ECMO centres are adhering to this recommendation with no more than five access occasions per day for over half of the respondents. Conversely, over one-quarter of respondents reported administering intermittent drug or electrolyte boluses via the circuit whereas only the administration of continuous infusions is recommended to minimise breaking circuit sterility[22]. This may reflect a shortage of access sites in these critically ill patients but using the ECMO circuit as an intermittent vascular access port is against current practice recommendations and should not occur.
Numerous types of antiseptic preparations were reported for cannula insertion and maintenance, and for accessing the circuit. Alcohol-based CHG was most commonly reported as the preferred antiseptic for both cannula insertion and maintenance, whilst its use was slightly less than alcohol swabs for circuit access. Alcohol was also used to disinfect access ports by the majority of centres surveyed by Glater-Welt et al.[21]. This practice conflicts with ELSOs recommendation to use CHG preparations for circuit antisepsis rather than alcohol or povidone iodine[22]. It is unclear from ELSO whether CHG should be contained within an alcohol-based or aqueous preparation. The prevalent use of CHG in alcohol identified in this survey aligns with guidelines relevant to central lines[25–27]. This combined preparation has proven more effective than either CHG or alcohol used alone due to both immediate and sustained disinfecting action[28]. However, further data regarding skin antisepsis regimens including the influence of aqueous or alcohol-based solutions on infection outcomes is needed[29].
Aqueous-based CHG preparation was identified as one of the top five preferred line antisepsis agents in this survey. Its use appears to correspond with anecdotal evidence from centres whose local policy dictates alcohol avoidance due to the potential to cause damage to plastics used in ECMO cannulas and circuit componentry. However, there is little evidence to corroborate this. Past reports have implicated acetone contained within the solution as the destructive agent, not alcohol alone[30]. Investigation into the effects of alcohol in this context would clarify this practice discrepancy.
Respondent centres identified the most common circuit connection on ports or stopcocks is via a luer-lock (‘screw-in’) method whereas the recommendation is for the use of needleless valves which are considered more reliably decontaminated with disinfecting solutions[22]. Evidence regarding the effectiveness of needleless connectors in infection prevention is unclear. They appear to reduce microbial colonisation and catheter-related BSI, but the mechanical type needleless connectors may increase infection risk[25, 26]. Evidence is based on comparisons between needleless connectors and stopcocks or caps used for intravascular devices and does not account for extracorporeal circuit ports. Interestingly, a small number of centres reported using passive disinfection, usually containing alcohol, on circuit ports. Disinfection caps have been shown to reduce intraluminal contamination and central catheter-related BSI rates[31] and thus have potential as a practicable disinfection strategy prior to making connections to the ECMO line ports. Research on the use of needleless connectors and disinfection caps in the ECMO context is needed.
Visual cannula site inspection, followed by swab and culture, was the most commonly reported local infection surveillance method. Therefore, it is important that any dressing placed over the insertion site is transparent to allow for continual inspection. Data on ECMO CRI and colonisation is limited and mostly derived from single-centre cohorts[1, 17, 18, 32] but is nonetheless of increasing concern as evidence evolves. Most recently Allou et al.[1] found that CRIs were frequent in patients supported by peripheral ECMO with 17.7% of patients developing this complication. Importantly it is likely the magnitude of the problem is not fully appreciated: CRIs may be under-reported in the absence of a consensus definition[15] and thus this complication may be under-recognised in clinical practice and less reliably reported in the evidence base. Patients on ECMO generally have multiple invasive lines in addition to the large-bore ECMO lines which increases the number of entry portals for intraluminal contaminants and constitutes an increased risk for BSI among other factors[6, 33]. Routine changeout of ECMO lines is unfeasible due to the high risk of life-threatening complications and patient cardiorespiratory instability precludes this as an option when potential or proven infections occur. Careful adherence to preventive measures and vigilant patient monitoring is vital to avoid this eventuality. There is a clear need for further research as there are no guidelines for CRI diagnosis and management at present.
Dressing and securement of peripheral ECMO lines
Dressing changes most commonly occurred either when necessary or between one to three days frequency, followed by daily. Glater-Welt et al.[21] similarly reported that ‘when necessary’ dressing changes were carried out by most of their surveyed centres. Transparent dressing changes for central catheters are recommended every seven days or sooner if dressing adherence or soiling is an issue[26]. The need for frequent dressing changes reported for ECMO patients may reflect cannulation site bleeding issues, however, dressing disruption is a major risk factor for catheter-related infections and should be avoided where possible[34]. The ELSO general guidelines[20] do not specify a frequency for cannula site antisepsis except to recommend that sites be cleaned ‘frequently’. The development of a cannula dressing regimen with replacement frequencies specific to ECMO is needed.
CHG-impregnated dressing products have been shown to reduce central catheter-related BSI and colonisation compared with standard transparent dressings[35], appear to be cost effective in adult intensive care patients[36, 37], and are recommended for use in patients over the age of two months[27]. Our finding that CHG dressing products, including disks and integrated gel dressings, are used by more than half the countries surveyed suggests uptake of this evidence in ECMO practice. Use of an integrated transparent CHG dressing product is more practical in ECMO because, unlike disks, this dressing type allows visualisation of the cannula site and eliminates the need for two products. Data on clinical efficacy at reducing infection as well as safety and cost-effectiveness is needed to justify the routine use of CHG dressings in ECMO.
In compliance with ELSO recommendations, the securement of peripheral cannulas to the skin in at least two locations to prevent inadvertent decannulation appears common practice[20], as does suturing at cannulation sites, yet the concerning reports of adverse patient events associated with cannula dislodgement highlights the need to clearly define optimal line securement. We identified inadequate securement as the top cause of adverse patient outcomes resulting from cannula malposition or dislodgement. Many of the other stated reasons (i.e. during ambulation, transport and patient cares) may also be attributed to suboptimal dressing and securement techniques. Maintaining line security throughout the duration of ECMO is an absolute patient safety priority to avoid decannulation-related life-threatening emergencies[20]. The present trend towards less sedation and early mobilisation in certain patient groups[38] further underlines this imperative. Our respondents reported numerous securement methods likely indicative of the heterogeneity of patient populations, cannulation methods, circuit configurations, and models of care practiced. Methods included various commercial products (i.e. ‘Grip-Lok’ style devices, Foley catheter devices, colostomy ring patches) not designed and/or marketed specifically for ECMO use. Some methods may be risky if used in novel ways outside of manufacturer recommendations[39]. Rigorous testing of available products or the development of ECMO-specific devices would be beneficial to ensure they allow for safe and robust fixation of large ECMO lines.
Cannula adverse events were higher in high-volume adult centres and no difference was identified in adverse events rates between centres with less than two years of ECMO provision and those with over 15 years. Reports of more cannula malposition or dislodgement in centres which always suture at the cannula insertion site compared with sites who never suture did not meet the a priori statistical significance level but is of clinical significance. The finding is somewhat counterintuitive as suturing would be expected to enhance securement of the ECMO cannula. This may indicate greater vigilance toward line monitoring and management in centres that do not routinely suture cannulas because this practice is thought to present a greater risk for the occurrence of adverse events. It must be noted though that reporting and recall bias may have affected the reporting of these adverse events.
Clinical practice guidelines for ECMO lines
There is presently a lack of high-quality evidence to guide ECMO line management practices. Recommendations from the ELSO[20, 22] are largely derived from consensus guidelines applicable to smaller intravascular devices[25, 26] and lack ECMO-specific data. Therefore, currently available evidence may be insufficient for the larger specialised lines used in ECMO. This is supported by the belief of most ECMO respondents that an international evidence-based guideline is needed to inform bedside practices in this regard.
Limitations
Due to the cross-sectional design our findings may not represent all practices in the ECMO community. Similarly, participants from the United States contributed almost half of the responses but this was expected given that the largest number of ECMO centres is present in this location. However, our survey had a good geographic distribution with responses from 45 of the 51 countries surveyed. Findings may also be limited by issues inherent with self-reported data including recall bias; social desirability bias; and question interpretation by respondents whose first language is not English. Our response rate is based on individual respondent numbers. It was not possible to calculate a per centre rate as we chose not to identify participant or institutional details, or record IP addresses for anonymity. Therefore, despite explicit instructions requesting one participant per centre, it is evident that some double enrolments occurred in this survey. We chose to focus on the pragmatic, routine, hands-on line practices undertaken at the bedside; in the interests of length our survey did not cover in-depth all line management aspects or issues (i.e. central cannulas, cannula bleeding, surveillance cultures, antimicrobial use, and infection rates), some of which have been previously surveyed[21, 40]. Nonetheless, our study is the first to comprehensively examine line dressing, securement and access practices to our knowledge.
Conclusions
Variable practices regarding ECMO line management exist worldwide based on recommendations lacking specific ECMO content and context. This survey identified evidence gaps regarding several aspects of patient management, specifically antiseptic agents, dressing products and regimens, securement methods and needleless valves. The vulnerability of ECMO patients for significant morbidity and mortality associated with line-related complications and ECMOs expanding use worldwide warrants targeted research to develop best practice guidelines for clinicians managing ECMO lines.
Supporting information
S1 Appendix [pdf]
Questionnaire tool.
S1 Dataset [xlsx]
Minimal anonymized data set.
Zdroje
1. Allou N, Pinto HL, Persichini R, Bouchet B, Braunberger E, Lugagne N, et al. Cannula-related infection in patients supported by peripheral ECMO: Clinical and microbiological characteristics. ASAIO Journal. 2019;65(2):180–6. doi: 10.1097/MAT.0000000000000771 29517513
2. Conrad SA, Grier LR, Scott LK, Green R, Jordan M. Percutaneous cannulation for extracorporeal membrane oxygenation by intensivists: A retrospective single-institution case series. Critical Care Medicine. 2015;43(5):1010–5. doi: 10.1097/CCM.0000000000000883 25746749
3. Butt W, Heard M, Peek GJ. Clinical management of the extracorporeal membrane oxygenation circuit. Pediatric Critical Care Medicine. 2013;14(5):S13–S9.
4. Belohlavek J, Rohn V, Tosovsky J, Kunstyr J, Semrad M, Horak J, et al. A review of a newly established ECMO program in a university affiliated cardiac center. Journal of Cardiovascular Surgery. 2011;52(3):445–51. 21577197
5. Lindskov C, Jensen RH, Sprogoe P, Klaaborg E, Kirkegaard H, Severinsen IK, et al. Extracorporeal membrane oxygenation in adult patients with severe acute respiratory failure. Acta Anaesthesiologica Scandinavica. 2013;57(3):303–11. doi: 10.1111/aas.12050 23278552
6. Aubron C, Cheng AC, Pilcher D, Leong T, Magrin G, Cooper DJ, et al. Infections acquired by adults who receive extracorporeal membrane oxygenation: Risk factors and outcome. Infection Control and Hospital Epidemiology. 2013;34(1):24–30. doi: 10.1086/668439 23221189
7. Bizzarro MJ, Conrad SA, Kaufman DA, Rycus P. Infections acquired during extracorporeal membrane oxygenation in neonates, children, and adults. Pediatric Critical Care Medicine. 2011;12(3):277–81. doi: 10.1097/PCC.0b013e3181e28894 20495508
8. Burket JS, Bartlett RH, Vender Hyde K, Chenoweth CE. Nosocomial infections in adult patients undergoing extracorporeal membrane oxygenation. Clinical Infectious Diseases. 1999;28(4):828–33. doi: 10.1086/515200 10825046
9. Hsu M-S, Chiu K-M, Huang Y-T, Kao K-L, Chu S-H, Liao C-H. Risk factors for nosocomial infection during extracorporeal membrane oxygenation. Journal of Hospital Infection. 2009;73(3):210–6. doi: 10.1016/j.jhin.2009.07.016 19782430
10. Sun H-Y, Ko W-J, Tsai P-R, Sun C-C, Chang Y-Y, Lee C-W, et al. Infections occurring during extracorporeal membrane oxygenation use in adult patients. The Journal of Thoracic and Cardiovascular Surgery. 2010;140(5):1125–32. doi: 10.1016/j.jtcvs.2010.07.017 20708754
11. Douglass BH, Keenan AL, Purohit DM. Bacterial and fungal infection in neonates undergoing venoarterial extracorporeal membrane oxygenation: An analysis of the registry data of the extracorporeal life support organization. Artificial Organs. 1996;20(3):202–8. doi: 10.1111/j.1525-1594.1996.tb04428.x 8694690
12. Meyer DM, Jessen ME, Eberhart RC. Neonatal extracorporeal membrane oxygenation complicated by sepsis. Extracorporeal Life Support Organization. Tha Annals of Thoracic Surgery. 1995;59(4):975–80.
13. Montgomery VL, Strotman JM, Ross MP. Impact of multiple organ system dysfunction and nosocomial infections on survival of children treated with extracorporeal membrane oxygenation after heart surgery. Critical Care Medicine. 2000;28(2):526–31. doi: 10.1097/00003246-200002000-00040 10708195
14. Biffi S, Di Bella S, Scaravilli V, Peri AM, Grasselli G, Alagna L, et al. Infections during extracorporeal membrane oxygenation: Epidemiology, risk factors, pathogenesis and prevention. International Journal of Antimicrobial Agents. 2017;50(1):9–16. doi: 10.1016/j.ijantimicag.2017.02.025 28528989
15. Messika J, Schmidt M, Aubry A, Combes A, Ricard J-D. Extracorporeal membrane oxygenation-associated infections: Carefully consider cannula infection. Critical Care Medicine. 2018;46(2):e171–e2.
16. Australia and New Zealand Extracorporeal Membrane Oxygenation (ANZ ECMO) Investigators, Davies A, Jones D, Bailey M, Beca J, Bellomo R, et al. Extracorporeal membrane oxygenation for 2009 influenza A(H1N1) acute respiratory respiratory distress syndrome. JAMA. 2009;302(17):1888–95. doi: 10.1001/jama.2009.1535 19822628
17. Combes A, Leprince P, Luyt CE, Bonnet N, Trouillet J-L, Leger P, et al. Outcomes and long-term quality of life of patients supported by extracorporeal membrane oxygenation for refractory cardiogenic shock. Critical Care Medicine. 2008;36(5):1404–11. doi: 10.1097/CCM.0b013e31816f7cf7 18434909
18. Schmidt M, Bréchot N, Hariri S, Guiguet M, Luyt CE, Makri R, et al. Nosocomial infections in adult cardiogenic shock patients supported by venoarterial extracorporeal membrane oxygenation. Clinical Infectious Diseases. 2012;55(12):1633–41. doi: 10.1093/cid/cis783 22990851
19. Hahne K, Horstmann C, Fischer D, Kock R, Peters G, Lebiedz P. Cannula-related infection in adult medical intensive care unit patients undergoing extracorporeal life support and extracorporeal membrane oxygenation. Journal of Hospital Infection. 2015;91(4):372–3. doi: 10.1016/j.jhin.2015.08.022 26518270
20. Extracorporeal Life Support Organization. ELSO guidelines for cardiopulmonary extracorporeal life support, Version 1.4. Ann Arbor, MI, USA: Extracorporeal Life Support Organisation; 2017 [cited 2018 Mar 19]. Available from: https://www.elso.org/Resources/Guidelines.aspx.
21. Glater-Welt LB, Schneider JB, Zinger MM, Rosen L, Sweberg TM. Nosocomial bloodstream infections in patients receiving extracorporeal life support: Variability in prevention practices: a survey of the extracorporeal life support organization members. Journal of Intensive Care Medicine. 2016;31(10):654–9. doi: 10.1177/0885066615571540 25670726
22. Extracorporeal Life Support Organization Infectious Disease Task Force. Infection control and extracorporeal life support; 2012 [cited 2018 Mar 19]. Available from: https://www.elso.org/AboutUs/TaskForces/InfectiousDiseaseTaskForce.aspx.
23. Gwet KL. Computing inter-rater realiability and its variance in the presence of high agreement. British Journal of Mathematical and Statistical Psychology. 2008;61(Pt 1):29–48. doi: 10.1348/000711006X126600 18482474
24. Klein D. "KAPPAETC: stata module to evaluate interrater agreement", Statistical Software Components S458283. Boston College Department of Economics; 2016 [updated 2018 Feb 1; cited 2018 Sep 5]. Available from: https://ideas.repec.org/c/boc/bocode/s458283.html.
25. O'Grady NP, Alexander M, Burns LA, Dellinger EP, Garland J, Heard SO, et al. Guidelines for the prevention of intravascular catheter-related infections. American Journal of Infection Control. 2011;39(4 Suppl 1):S1–34.
26. Loveday HP, Wilson JA, Pratt RJ, Golsorkhi M, Tingle A, Bak A, et al. Epic3: national evidence-based guidelines for preventing healthcare-associated infections in NHS hospitals in England. Journal of Hospital Infection. 2014;86(Suppl 1):S1–S70.
27. Yokoe DS, Anderson DJ, Berenholtz SM, Calfee DP, Dubberke ER, Ellingson KD, et al. A compendium of strategies to prevent healthcare-associated infections in acute care hospitals: 2014 updates. American Journal of Infection Control. 2014;42(8):820–8. doi: 10.1016/j.ajic.2014.07.002 25087135
28. Moreau NL, Flynn J. Disinfection of needleless connector hubs: clinical evidence systematic review. Nursing Research and Practice. 2015. doi: 10.1155/2015/796762 26075093. PubMed Central PMCID: PMC4446481.
29. Lai NM, Lai NA, O'Riordan E, Chaiyakunapruk N, Taylor JE, Tan K. Skin antisepsis for reducing central venous catheter-related infections. Cochrane Database of Systematic Reviews. 2016. doi: 10.1002/14651858.CD010140.pub2 27410189. PubMed Central PMCID: PMC6457952.
30. Niles SD, Ploessl J, Sutton PG, Steinberg JB. Oxygenator failure due to contact with bathing alcohol: A case report. Journal of Extra-corporeal Technology. 1992;24(2):69–71.
31. Gorski L, Hadaway L, Hagle ME, McGoldrick M, Orr M, Doellman D. Infusion therapy standards of practice. Journal of Infusion Nursing. 2016;29(1S):S1–S159.
32. Thomas G, Hraiech S, Cassir N, Lehingue S, Rambaud R, Wiramus S, et al. Venovenous extracorporeal membrane oxygenation devices-related complications and infections. Annals of Intensive Care. 2017;7:111. doi: 10.1186/s13613-017-0335-9 29116464; PubMed Central PMCID: PMC5676570.
33. Na SJ, Chung CR, Choi HJ, Cho YH, Yang JH, Suh GY, et al. Bloodstream infection in patients on venovenous extracorporeal membrane oxygenation for respiratory failure. Infection Control and Hospital Epidemiology. 2018;39(7):871–4. doi: 10.1017/ice.2018.90 29733005
34. Timsit JF, Mimoz O, Mourvillier B, Souweine B, Garrouste-Orgeas M, Alfandari S, et al. Randomized controlled trial of chlorhexidine dressing and highly adhesive dressing for preventing catheter-related infections in critically ill adults. American Journal of Respiratory and Critical Care Medicine. 2012;186(12):1272–8. doi: 10.1164/rccm.201206-1038OC 23043083
35. Ullman AJ, Cooke ML, Mitchell M, Lin F, New K, Long DA, et al. Dressing and securement for central venous access devices (CVADs): A Cochrane systematic review. International Journal of Nursing Studies. 2016;59:177–96. doi: 10.1016/j.ijnurstu.2016.04.003 27222463.
36. Jenks M, Craig J, Green W, Hewitt N, Arber M, Sims A. Tegaderm CHG IV securement dressing for central venous and arterial catheter insertion sites: A NICE medical technology guidance. Applied Health Economics and Health Policy. 2016;14(2):135–49. doi: 10.1007/s40258-015-0202-5 26458938
37. Schwebel C, Lucet JC, Vesin A, Arrault X, Calvino-Gunther S, Bouadma L, et al. Economic evaluation of chlorhexidine-impregnated sponges for preventing catheter-related infections in critically ill adults in the dressing study. Critical Care Medicine. 2012;40(1):11–7. doi: 10.1097/CCM.0b013e31822f0604 21926570
38. Combes A, Brodie D, Chen YS, Fan E, Henriques JPS, Hodgson C, et al. The ICM research agenda on extracorporeal life support. Intensive Care Med. 2017;43(9):1306–18. doi: 10.1007/s00134-017-4803-3 28470346
39. Inwood I. An exploration of the past, present and future of catheter securement. British Journal of Nursing. 2014;23(8):S26–S7. doi: 10.12968/bjon.2014.23.Sup8.S26 24763271
40. Kao LS, Fleming GM, Escamilla RJ, Lew DF, Lally KP. Antimicrobial prophylaxis and infection surveillance in extracorporeal membrane oxygenation patients: A multi-institutional survey of practice patterns. ASAIO Journal. 2011;57(3):231–8. doi: 10.1097/MAT.0b013e31820d19ab 21317768
Článok vyšiel v časopise
PLOS One
2019 Číslo 12
- Metamizol jako analgetikum první volby: kdy, pro koho, jak a proč?
- Nejasný stín na plicích – kazuistika
- Masturbační chování žen v ČR − dotazníková studie
- Úspěšná resuscitativní thorakotomie v přednemocniční neodkladné péči
- Fixní kombinace paracetamol/kodein nabízí synergické analgetické účinky
Najčítanejšie v tomto čísle
- Methylsulfonylmethane increases osteogenesis and regulates the mineralization of the matrix by transglutaminase 2 in SHED cells
- Oregano powder reduces Streptococcus and increases SCFA concentration in a mixed bacterial culture assay
- The characteristic of patulous eustachian tube patients diagnosed by the JOS diagnostic criteria
- Parametric CAD modeling for open source scientific hardware: Comparing OpenSCAD and FreeCAD Python scripts