Comparison of quality control methods for automated diffusion tensor imaging analysis pipelines


Authors: Seyyed M. H. Haddad aff001;  Christopher J. M. Scott aff002;  Miracle Ozzoude aff002;  Melissa F. Holmes aff002;  Stephen R. Arnott aff003;  Nuwan D. Nanayakkara aff001;  Joel Ramirez aff002;  Sandra E. Black aff002;  Dar Dowlatshahi aff005;  Stephen C. Strother aff003;  Richard H. Swartz aff004;  Sean Symons aff008;  Manuel Montero-Odasso aff009;  ;  Robert Bartha aff001
Authors place of work: Centre for Functional and Metabolic Mapping, Robarts Research Institute, University of Western Ontario, London, Ontario, Canada aff001;  L.C. Campbell Cognitive Neurology Research Unit, Hurvitz Brain Sciences Research Program, Sunnybrook Research Institute, University of Toronto, Toronto, Ontario, Canada aff002;  Rotman Research Institute, Baycrest Centre, Toronto, Ontario, Canada aff003;  Department of Medicine, Division of Neurology, Sunnybrook Health Sciences Centre, and University of Toronto, Toronto, Ontario, Canada aff004;  Ottawa Hospital Research Institute, Ottawa, Ontario, Canada aff005;  Department of Medical Biophysics, University of Toronto, Toronto, Ontario, Canada aff006;  Sunnybrook Health Sciences Centre, University of Toronto, Stroke Research Program, Toronto, Ontario, Canada aff007;  Department of Medical Imaging, Sunnybrook Health Sciences Centre, Toronto, Ontario, Canada aff008;  Department of Medicine, Division of Geriatric Medicine, Parkwood Hospital, University of Western Ontario, London, Ontario, Canada aff009;  Department of Medical Biophysics, University of Western Ontario, London, Ontario, Canada aff010
Published in the journal: PLoS ONE 14(12)
Category: Research Article
doi: https://doi.org/10.1371/journal.pone.0226715

Summary

The processing of brain diffusion tensor imaging (DTI) data for large cohort studies requires fully automatic pipelines to perform quality control (QC) and artifact/outlier removal procedures on the raw DTI data prior to calculation of diffusion parameters. In this study, three automatic DTI processing pipelines, each complying with the general ENIGMA framework, were designed by uniquely combining multiple image processing software tools. Different QC procedures based on the RESTORE algorithm, the DTIPrep protocol, and a combination of both methods were compared using simulated ground truth and artifact containing DTI datasets modeling eddy current induced distortions, various levels of motion artifacts, and thermal noise. Variability was also examined in 20 DTI datasets acquired in subjects with vascular cognitive impairment (VCI) from the multi-site Ontario Neurodegenerative Disease Research Initiative (ONDRI). The mean fractional anisotropy (FA), mean diffusivity (MD), axial diffusivity (AD), and radial diffusivity (RD) were calculated in global brain grey matter (GM) and white matter (WM) regions. For the simulated DTI datasets, the measure used to evaluate the performance of the pipelines was the normalized difference between the mean DTI metrics measured in GM and WM regions and the corresponding ground truth DTI value. The performance of the proposed pipelines was very similar, particularly in FA measurements. However, the pipeline based on the RESTORE algorithm was the most accurate when analyzing the artifact containing DTI datasets. The pipeline that combined the DTIPrep protocol and the RESTORE algorithm produced the lowest standard deviation in FA measurements in normal appearing WM across subjects. We concluded that this pipeline was the most robust and is preferred for automated analysis of multisite brain DTI data.

Keywords:

Diffusion tensor imaging – magnetic resonance imaging – Data acquisition – Computational pipelines – Data processing – Image processing – Diffusion weighted imaging – Graphics pipelines

Introduction

Diffusion tensor imaging (DTI) is a well-established magnetic resonance imaging (MRI) technique, sensitive to the microstructural organization of cerebral tissue constituents [1,2]. This technique can be used to characterize healthy cerebral tissue microstructure as well as diverse subtle pathological and age-related alterations within the brain [1,3]. Diffusion tensor imaging can also be used to characterize primary water diffusion paths along axonal fibers to study structural connectivity and white matter (WM) integrity [4].

To effectively interpret brain DTI data a number of consecutive image processing steps must be performed including data conversions, artifact removal (such as eddy current (EC), motion, and gradient distortion corrections), and tensor fitting before diffusion metrics are calculated [511]. Recently, quality control (QC) procedures to remove outliers (e.g. distorted gradient volumes) from the analysis, modify voxel-wise diffusion tensor components, and harmonize data have also been identified as necessary steps in DTI analysis pipelines [5,810,1216]. The latter step, is particularly relevant when analyzing large multisite and/or multiscanner DTI datasets [9,1517], including those from the UK Biobank project [5,1820] and the Ontario Neurodegenerative Disease Research Initiative (ONDRI) [21].

Implementing such complicated [11] processing requires automation to reduce unconscious user errors [22] and increase processing speed for large datasets [9,1517]. Concerted efforts have been directed towards implementing fully automatic image processing pipelines for the analysis of brain DTI data [510,13]. For example, the Enhanced Neuroimaging Genetics though Meta-Analysis (ENIGMA) Consortium, a collaborative network of researchers working in the areas of neuroscience, genetics, and medicine, have analyzed neuroimaging data from thousands of subjects [9,23,24]. The ENIGMA DTI working group has created a well-known and widely accepted framework for region of interest (ROI) and tract-based analyses, and harmonization of DTI data from multiple sites [9,24]. In particular, the ENIGMA DTI working group focus has been on the analysis of high resolution multisite fractional anisotropy (FA) maps from healthy subjects to study the influence of genetic factors on the intersubject variabilities of WM microstructure [23,24]. Similarly, the Human Connectome Project (HCP), a consortium of institutions across the Unites States and Europe, aims to delineate brain connectivity and WM networks based on brain DTI analysis and tractography in a large numbers of healthy individuals [10,25].

Recently minimal preprocessing pipelines for automatic analysis of the HCP DTI data and other HCP MR imaging modalities were introduced [10]. In the HCP DTI preprocessing pipeline, essential image processing steps such as artifact (EC and gradient distortion) removal, surface generation, cross-modal linear registration, and transformation to a standard space were included [10]. This pipeline was designed to work with high quality DTI data produced by the HCP [10] and accordingly some QC procedures such as outlier detection and the removal of slice-wise and gradient-wise inconsistencies which are typically needed for conventional DTI data [13] were not included.

In the current study, we present, validate, and compare three fully automatic ENIGMA-based pipelines for processing brain DTI data by effectively connecting multiple image processing tools. We tested the performance of different automatic QC procedures as a supplementary image processing step added to the general ENIGMA recommended framework. For this purpose, two major types of QC procedures were considered. The first one is DTIPrep [13,26], which fully rejects diffusion weighted volumes along gradient directions significantly affected by gradient distortion artifacts [13]. The second one is Robust Estimation of Tensors by Outlier Rejection (RESTORE) [27]. RESTORE iteratively detects local (voxel-wise) outliers in the diffusion volumes and removes them from the final tensor estimation procedure [27]. Using these QC procedures, we devised three pipelines complying with the general ENIGMA framework. The first one used RESTORE as a QC procedure, the second one used the DTIPrep protocol, and the third one used a combination of both. The performance of the pipelines was validated and evaluated using “ground truth” and “artifactual” simulated diffusion weighted images (DWIs) generated based on the frameworks introduced in [28,29]. Pipelines were also applied to multi-site diffusion data from the ONDRI study to assess variance in normal tissue. We hypothesized that the third pipeline, which combined both QC procedures would produce optimal measurement precision by generating a smaller range of DTI parameter values in normal appearing WM.

Materials and methods

Simulated DWI data

POSSUM (Physics-Oriented Simulated Scanner for Understanding MRI) is an FSL toolbox that simulates MRI data by solving the Bloch and Maxwell equations. The toolbox produces realistic MR images and artifacts [28]. The simulated DWI data in this study were generated by an extension to POSSUM [30,31] called DW-POSSUM [28]. DW-POSSUM [29] effectively simulates DWI data and related artifacts such as eddy current induced artifacts. Here we adopted two major types of DWI datasets from [29,32]: a ground truth DWI dataset and four corresponding datasets with motion artefact, eddy current artefact, and thermal noise (Rician noise). Specifically, one dataset contained a large amount of motion at SNR = 20 (LM-20), a second dataset contained a large amount of motion at SNR = 40 (LM-40), a third dataset contained a small amount of motion at SNR = 20 (SM-20), and a fourth dataset contained a small amount of motion at SNR = 40 (LM-40). Motion artifacts were added to the simulated datasets by applying various degrees of rigid rotations to the geometric object (head), before the MR simulation began, in order to model instantaneous between-volume motions [28,29]. Detailed profiles of the head rotations used to generate motion artifacts in both the large and small motion scenarios were provide in reference [29]. POSSUM required two major inputs to simulate DWI datasets. First a realistic brain object was provided by a full-brain segmentation of high resolution T1- and T2-weighted images (matrix size: 260×311×260 and voxel size: 0.7 mm) of a single subject from the WU-Minn HCP dataset [25]. These images were segmented using the FSL Fast toolbox [33]. Second, a diffusion-weighted pulse sequence was simulated by a voxel-wise spherical harmonic fit (order n = 8) [34] to the b = 700 s/mm2 shell considering 32 gradient directions plus four b = 0 images [28,29]. The image acquisition was simulated using an echo-planar imaging (EPI) pulse sequence with TE = 109 ms and TR = 7500 ms and a matrix size of 72×68×55 (isotropic voxel size of 2.5 mm).

ONDRI participants

We included 20 participants with VCI aged 60–85 years (40% female) available in the ONDRI database. ONDRI is a multi-site longitudinal research study of diverse neurodegenerative diseases including VCI [21,35]. ONDRI includes neuroimaging and genomics in addition to assessments of cognition as well as language, speech, gait, retinal imaging, and eye tracking [21,35]. The study was approved by the Ethics Review Boards at all participating institutions [21].

Magnetic resonance imaging data acquisition

The acquisition of neuroimaging data for ONDRI participants has been previously described and was consistent with the Canadian Dementia Imaging Protocol (CDIP) [21]. Briefly, T1-weighted, T2-weighted, fluid-attenuated inversion recovery (FLAIR), T2*-weighted, positron density (PD)-weighted image, resting state functional MRI (fMRI), and diffusion MRI (dMRI) images were acquired from 11 different 3.0T scanners across Ontario. Scanner specific details are provided in Table 1.

Tab. 1. DTI data acquisition parameters for the different MRI scanners utilized in the current study.
DTI data acquisition parameters for the different MRI scanners utilized in the current study.

For the current study, dMRI data were available and incorporated from seven different scanners/sites. The DTI processing pipelines utilized T1-weighted (voxel dimensions of 1.0×1.0×1.0 mm3), T2-weighted images (0.9375×0.9375×3.0 mm3), and dMRI data. The details of the acquisition parameters and pulse sequences for the DTI data across the seven scanners used for analysis are provided in Table 1. All the ONDRI DTI data used in this study were produced with 30 different gradient directions (with b = 1000 s/mm) with voxel dimensions of 2.0×2.0×2.0 mm3. The DTI data produced on Siemens and Philips scanners were acquired with one b0 volume, while in the case of GE scanners, three b0 volumes were acquired. On the Siemens and Philips scanners, two additional b0 images were acquired with phase encoding reversed. These were utilized for EPI distortion corrections in the processing pipelines. The simulated DWI data used in this study were comparable to the ONDRI data in terms of the number of the gradient directions as well as b-value.

Segmentation of cerebral tissues and vascular lesions

The procedures used to segment regions of interest have been detailed previously (Semi-Automatic Brain Region Extraction (SABRE) [36], Lesion Explorer [37,38], and Fuzzy Lesion EXtractor (FLEX) [39]), including a scan-rescan reliability analysis [40]. Briefly, interleaved proton density (PD), T2-weighted images, and FLAIR images were co-registered to the T1-weighted image, and a PD-T2 based mask was automatically generated and manually edited. Using this mask, the T1-weighted image was segmented using a multi-feature histogram method [41] to generate GM, WM, and cerebrospinal fluid (CSF) regions, and ventricular CSF (vCSF) was manually relabeled. WM hyperintensities (WMHs) and lacunae were also automatically identified using FLEX (FLAIR-based) and Lesion Explorer along with manual edits. Each lesion type was further subdivided into a region-based class (periventricular or deep) by an automated algorithm [3739]. Lesion Explorer was also used to capture enlarged perivascular spaces (PVS) [38,42], and cortical strokes were manually traced. A combination of these tissue segmentation methods produced ten different tissue classes including normal appearing WM (NAWM), which was used in the current study to evaluate DTI processing pipeline performance.

Ground truth DTI metrics

The ground truth DTI metrics were calculated by rigidly registering the (artifact-free) simulated DTI data to the high resolution T1-weighted image and then tensor fitting to calculate DTI values at each voxel. Registering the DTI data to the T1 space is part of each pipeline as recommended by ENIGMA framework.

Automated DTI processing pipelines

Three different fully automated pipelines were designed for processing and analysis of the brain DWIs. All the pipelines complied with the general ENIGMA DTI framework, a well-established methodology for processing diffusion data [9,24]. In the proposed pipelines we effectively used and optimized seven different image processing software packages including 3D Slicer [43], FSL [44], Ants [45], Camino [46], DTIPrep [13], MRIcron [47], and DTI-TK [48]. In all pipelines, automated QC procedures were added to the ENIGMA DTI framework, allowing detection of outlier data for removal prior to calculation of the diffusion tensor. These QC procedures are essential in multisite studies, where data is acquired on typical scanners available in healthcare centers [21]. All the proposed pipelines were automated using Bash scripts that are executable on Linux-based operating systems. All analyses were performed on a commodity computer work station with an Intel Core i7-6800k @3.4 GHz × 12 CPU, 16 GB RAM, GeForce GTX 1050 Graphics card, and a Linux Ubuntu 64-bit operating system.

The proposed pipelines are illustrated in Fig 1. As we used three different MR modalities, i.e. T1-weighted, T2-weighted, and diffusion images in all the proposed pipelines, we describe the subpipelines related to the processing of each MR modality as shown in Fig 1. The three proposed pipelines differ only in the DTI processing subpipelines.

Fig. 1. Schematic representation of the three pipelines utilized for brain DTI data processing.
Schematic representation of the three pipelines utilized for brain DTI data processing.
The T1- and T2-weighted imaging processing subpipelines are common to all DTI processing pipelines. The blue, red, and purple arrows depict the 1st, 2nd, and 3rd DTI processing pipelines, respectively. T1 = T1-weighted image, T2 = T2-weighted image, b0 = b0 image.

Pipeline #1 based on robust tensor fitting

T1-weighted image processing subpipeline. The DTI data were registered to the native T1-weighted image space. This was done so that subject specific tissue segmentations could be applied to the DTI scalar maps to obtain ROI based DTI scalar metrics. Moreover, the T1-weighted image was used as the target of the nonlinear registration of the DWI data in the proposed DTI pipelines, to improve the susceptibility-induced EPI distortions on the diffusion data. The following image processing procedures were performed on the T1 images (Fig 1):

  • DICOM to NIFTI conversion. The T1-weighted image was converted from DICOM [49] to 3D NIFTI [50] volumes so that image processing procedures could be applied to the data. Tools such as DWI Converter in 3D Slicer software v4.8.0 [51,52] or dcm2nii in MRIcron [53] can accomplish this image transformation. In the proposed pipelines we used DWI Converter in 3D Slicer (v4.8.0).

  • Gradient nonlinearity distortion correction. The NIFTI T1-weighted image was transformed to eliminate gradient nonlinearity artifacts from the image [54] using a procedure previously described [55]. Gradient distortions alter brain geometry [54] and may reduce the accuracy of tissue segmentation. This step was performed in an attempt to reduce site-specific image distortions to improve the precision of the quantitative morphometric analyses [54].

  • Bias field correction. Inhomogeneity of the magnetic field within the MR scanner is inevitable due to two major factors: 1) inherent inhomogeneity of the MRI magnet, and 2) inhomogeneity of the magnetic properties of the subject [33]. This inhomogeneity leads to low spatial frequency intensity (gray value) alteration across the image known as a bias field [33]. The bias field artifacts can be eliminated from the T1-weighted image using tools such as FAST in FSL (v5.0) [33] or N4ITK in 3D Slicer (v4.8.0) [56] using random field estimation models and expectation-maximization algorithms.

  • Skull stripping. The brain was extracted from the skull and other tissues surrounding the brain to increase processing efficiency and improve the performance of the registration of the DWI data and T2-weighted images to the T1-weighted images. There are many different algorithms and software tools to extract the brain from the T1-weighted structural images. In the current study, we used the robust brain extraction tool (ROBEX) in 3D Slicer (v4.8.0) [57] for skull stripping, which provided consistent high quality brain extraction in most cases tested. It is noteworthy that the performance and quality of the image extraction tools depend on the quality of the MR images.

T2-weighted image processing subpipeline. The T2-weighted MR images were utilized in all the proposed DTI processing pipelines to improve the quality of the nonlinear registration of the DWI data (mainly the b = 0 diffusion image) to the T1-weighted structural image. To the best of our knowledge, the nonlinear registration of DWI data to the T1-weighted structural image is challenging using common registration schemes [58,59]. This stems from the fact that both the T1-weighted images and the b = 0 diffusion images have different tissue intensity profiles for WM, GM, and CSF, which hinders the functionality of current similarity metrics (such as mutual information) [5860]. As the T2-weighted images have similar signal intensity characteristics to those found in the b = 0 image, the similarity metrics typically used in registration algorithms can more precisely assess the resemblance between these images (i.e. b = 0 and T2-weighted). Therefore, to improve the quality of this nonlinear registration, the b = 0 image was registered to a high quality non-EPI T2-weighted image as an intermediate stage for nonlinear registration of the b = 0 image to the T1-weighted structural image. To facilitate the nonlinear registration of the diffusion data to the T1-weighted image, both the b = 0 (diffusion data) and T2-weighted images were first linearly registered to the native T1-weighted structural image and then a nonlinear registration was performed between the T2-weighted and b = 0 images after both were linearly registered to the corresponding native T1-weighted image.

The T2-weighted image processing subpipeline was similar to the T1-weighted image processing subpipeline (Fig 1) except for two major differences: 1) gradient distortion correction was not applied to the T2-weighted image and 2) the T2-weighed image processing subpipeline incorporated a linear registration of the T2-weighted image to the T1-weighted structural image as explained above. To increase the accuracy of the linear registration of the T2-weighted image to the T1-weighted image, two major steps were considered in the T2-weighted image processing subpipeline. First, the T2-weighted image was linealy registered to the T1-weighted image using the well known linear registration tool FSL FLIRT [61,62]. Then, alignment was improved using a boundary-based registration (BBR) tool avaialble in FSL that was applied to the output of the first registration [63].

Diffusion weighted imaging processing subpipeline. The diffusion weighted imaging processing subpipeline involved a number of image processing procedures (Fig 1):

  • DICOM to Nearly Raw Raster Data (NRRD) conversion. We used the DWIConverter module in 3D Slicer (v4.8.0) to transform the DICOM data from the scanner to a 3D NRRD volume rather than NIFTI format. This step was performed because the Rician linear minimum mean square error (LMMSE) Image Filter Module in 3D Slicer used in the next stage of the subpipeline requires NRRD input images.

  • Denoising DWI volumes. DWI typically suffers from low SNR and is thus highly prone to noise artefacts [6466]. DTI is known to be affected by Rician noise [6466], which imposes a positive bias on the calculated voxelwise diffusion tensors [64]. Consequently, it is vital to reduce the level of Rician noise in the diffusion data. The level of the denoising must not oversmooth the data, which decreases image resolution and may reduce the ability to differentiate diverse types of cerebral tissues/lesions [67]. Therefore, we moderately denoised the DWI data using the Rician LMMSE Image Filter Module in 3D Slicer [66], with the iteration parameter of the filter set to 1. This denoising step in the DTI processing pipeline can be regarded as a QC procedure moderating the effect of the noise on the DWI data.

  • NRRD to NIFTI converter. After denoising, the NRRD data was converted to NIFTI format, for compatibility with most of brain DTI processing software packages including FSL and Camino. For this image conversion, we used the DWIConverter module in 3D Slicer (v4.8.0).

  • Estimation of the susceptibility-induced off resonance field. Off resonance spins arising from inhomogeneities of the static magnetic field, can produce artefacts in the DWI volumes that appear as unidirectional stretching or compression [68,69]. These off resonance spins occur due to magnetic susceptibility variations throughout the brain known as susceptibility-induced off resonance fields [68,69]. A systematic approach to eliminate the susceptibility-induced off resonance field from the EPI images is to collect either a magnetic field map or a second DWI dataset with reverse phase-encode (PE) direction during each specific diffusion acquisition [70]. Neither of these two options were implemented in the ONDRI DWI data acquisition, as greater scan time is needed, which is problematic for the ONDRI cohort suffering from various neurodegenerative conditions. Consequently, to minimize the effect of the susceptibility-induced artifacts on the ONDRI DWI data, we used a nonlinear registration to the T1-weighted structural image based on the recommendation made in the ENIGMA DTI framework (khaki dashed rectangle in Fig 1) [9,24]. Meanwhile, b = 0 images obtained on Siemens and Philips platforms as part of the DWI dataset were acquired with reverse PE direction, which was input to the well-known FSL top-up tool to estimate the susceptibility-induced off resonance field [68]. This estimated susceptibility-induced off resonance field in the absence of the DWI data with reverse PE direction was utilized as an input argument to the FSL eddy tool, to correct susceptibility-induced artefacts across the diffusion images [69]. Additional b = 0 images with reverse PE direction were not acquired on GE scanners, and hence this step was not performed (Fig 1).

  • EC-induced distortion and motion correction. Eddy currents induced in the magnet bore by fast gradient-switching during the diffusion weighted echo planar imaging sequence produces a time-dependent off resonance field [69]. In addition, head motion during long diffusion scans produces shifts in head position at different diffusion weightings [69]. Both artefacts are tackled in the FSL eddy tool by registering each individual gradient image to a model-free prediction of that image [69]. To increase the computational efficiency of the eddy and motion corrections procedure, we used FSL (v5.0) eddy_openmp, which is executable on multiple CPU cores using parallelization algorithms. The FSL eddy (step 5) and topup (step 4) tools used in the 1st DTI processing pipeline were adopted from FSL version 5.0.9. In that version correcting slice-to-volume movement and correcting susceptibility-by-movement interactions with eddy were not available and hence were not used. These options may be used to apply appropriate eddy/motion corrections in case of DTI data with severe motion artifacts.

  • EPI distortion correction. As indicated previously neither a magnetic field map nor DWI data with reverse PE direction were acquired. Hence, EPI distortion correction was achieved in the presented DTI processing pipeline using a nonlinear registration to the corresponding T1-weighted structural images [71,72]. Registration of the DWI data to the T1-weighted images as described in the T2–weighted images processing subpipeline section was performed using a two-stage registration scheme. First, the DWI b = 0 image was linearly registered to the T1-weighted structural image using the FSL (v5.0) BBR tool. Then, the SyN registration algorithm implemented in the ANTs software [73,74] was utilized for nonlinear registration of the b = 0 image to the T2-weighted image, which was itself already transferred to the native T1-weighted image space using a linear registration. Prior to the SyN nonlinear registration, the skull was stripped in the DWI data to improve the quality and computational efficiency of the nonlinear registration. Skull stripping was performed in the diffusion processing subpipeline using FSL BET2 [75].

  • Robust tensor fitting. In this step, an automatic QC procedure was performed on the DWI data. For this purpose, we used the RESTORE algorithm in the Camino software, which detects local voxelwise outliers in the DTI data and excludes them from the final tensor estimation procedure using iteratively reweighted least-square regression [27].

  • Calculate DTI scalar metrics. In this step, DTI scalar metrics including FA, mean diffusivity (MD), axial diffusivity (AD), and radial diffusivity (RD) maps were calculated throughout the brain. Diverse tools such as Camino dtfit [76,77], FSL (v5.0) dtifit [78,79], and DTI-TK TVtool [80] can be used to estimate voxelwise DTI metrics. In the 2nd pipeline, we used FSL (v5.0) dtifit while in the 1st and 3rd pipelines the DTI-TK TVtool was utilized.

  • Voxelwise QC procedure. At this final step the diffusion tensor in all the voxels throughout the brain was checked to obtain more reliable DTI metrics. To this end, the voxels whose diffusion tensors had at least one negative eigenvalue or whose AD, MD, or RD were higher than 5.5 were excluded from the ROI analysis [1].

Pipeline #2 based on the DTIPrep protocol

This pipeline is similar to Pipeline #1 except the QC procedure performed in step 7 of the DTI processing pipeline described above was removed and replaced by the DTIPrep QC protocol (v1.2.8) [13,26] after denoising (Fig 1).

DTIPrep QC protocol. This automatic QC protocol utilizes global rejection of the diffusion volumes affected by various DWI artefacts such as inter-slice brightness artefacts, venetian blind artefacts, eddy current distortions, and motion artefacts, beyond the threshold limits defined by the DTIPrep protocol [13]. During the DTIPrep QC procedure, all gradient volumes were linearly registered to the b = 0 image in the DWI data [13]. Moreover, in the 2nd pipeline, the topup, EC, and motion correction blocks considered in pipeline #1 (i.e. steps 4 and 5) were also skipped as they were implemented inside the DTIPrep automatic QC software package [13].

Pipeline #3 combining the DTIPrep protocol with robust tensor estimation

In the 3rd pipeline, both QC pipelines were applied (Fig 1) to detect outliers and remove them from brain DWI data.

Evaluation of the processing error of the pipelines

Each DTI processing pipeline introduces some error during the processing of the MR images. To evaluate this error all of the proposed pipelines were utilized to process the simulated ground truth DWI data. Note that the EPI distortion correction was not included in the processing of the ground truth DWI data as the datasets did not include any EPI-related artifacts. Moreover applying the denoising Rician LMMSE filter was also omitted because the dataset did not include any noise and applying this additional filter would have introduced unnecessary artifacts.

To evaluate the processing error in each pipeline, the difference of the mean value of each DTI metric, i.e. FA, MD, AD, and RD from the ground truth values were examined in both GM and WM regions. This difference was then normalized to the corresponding ground truth mean DTI metric value to calculate the percentage error associated with each DTI metric. To compare pipeline performance, the percentage errors associated with FA and MD calculations in eight GM and WM regions including hippocampus, thalamus, temporal lobe-GM, temporal lobe-WM, parietal lobe-GM, parietal lobe-WM, frontal lobe-GM, and frontal lobe-WM were examined. For a more precise voxelwise comparison, the FA and MD of each voxel in the WM region of the brain in the subject that was used to generate the ground-truth DTI data was examined. Voxels were excluded with very small FA or MD values (FA or MD<1e-5) that were unrealistic and likely due to inaccuracies associated with the WM region mask. FA and MD values produced by the three DTI processing pipelines were compared using ANOVA. The distribution of the FA and MD values as well as the ground-truth data were also examined. The Pearson correlation coefficients between the voxelwise FA values and MD values were also calculated between each pipeline.

Pipeline Evaluation based on the simulated artifactual DWI data

All of the proposed pipelines were also applied to four imperfect datasets: LM-20, LM-40, SM-20, and SM-40 [28,32]. Note that the EPI distortion correction was not applied because these datasets do not include EPI-related artifacts [28]. Similar to the previous section, the percentage error of the mean value of the DTI metrics from the ground truth value was measured in both GM and WM regions. To compare pipelines performance, the percentage errors associated with FA and MD calculations in eight GM and WM regions were examined.

Evaluation of the pipelines based on in vivo ONDRI DWI data

All the proposed pipelines along with an additional DTI processing pipeline that included no quality control steps, eddy current/motion correction, or artifact/noise removal but did apply linear and non-linear registration to the T1 space (pipeline-no QC), were applied to the 20 DWI datasets from the ONDRI VCI cohort to examine pipeline performance. FA and MD measurements were compared in NAWM using ANOVA followed by post hoc tests to identify differences between the pipelines.

Pipeline performance following denoising

To reduce Rician noise in a DWI dataset the Gradient Anisotropic Diffusion (GAD) filter [81] that is implemented in 3D Slicer was applied. This filter was tested on simulated the LM-20 dataset, which had the lowest quality among the tested datasets considering both the magnitude of motion artifacts and level of thermal noise. Parameters were chosen for the GAD filter (Conductance = 1, number of iterations = 5) to avoid blurring the edges of the image [82]. The percentage error of the DTI metrics were compared for all pipelines in WM regions between the LMMSE denoising filter (default filter, Fig 1) and the GAD denoising filter.

Results

Image processing

Ground truth dataset

All three pipelines successfully produced parametric maps of FA, MD, AD, and RD after processing the raw data without noise and artefacts that was comparable to the ground truth DTI metric maps obtained directly from ground truth data through tensor fitting without any further processing (Fig 2). The average run time for Pipeline #1 and Pipeline #3 was ~360 minutes, while Pipeline #2 took ~120 minutes to complete. Pipelines #1 and #3 likely required more time to process the ground truth data because the RESTORE algorithm consumed time (more iterations) for iterative reweighting to fulfill the goodness of fit criterion needed for successful robust tensor fitting.

Fig. 2.
(A) FA, (B) MD, (C) AD, and (D) RD maps produced from the raw ground truth DWI data by the three DTI processing pipelines and directly from the ground truth data. First, second, and third columns correspond to the 1st, 2nd, and 3rd DTI processing pipeline, respectively. The fourth column corresponds to the maps obtained directly from the ground truth data. (E) The corresponding T1-weighted structural image and (F) full-brain segmentation of cerebral tissues (WM, GM, and CSF) are also provided. The images were not interpolated.

Simulated data containing artefacts

All three pipelines successfully produced parametric maps of FA, MD, AD, and RD after processing the simulated DWI datasets containing artefacts: LM-20, LM-40, SM-20, and SM-40. The FA and MD maps corresponding to these datasets are shown in Figs 3 and 4, respectively. These maps are qualitatively similar to the ground truth FA (Figs 2a4) and MD (Figs 2b4) maps, showing similar patterns of contrast. The average run time per dataset was ~120 minutes for all pipelines.

Fig. 3. FA maps produced by the three proposed DTI processing pipelines from the DWI datasets containing artefacts.
FA maps produced by the three proposed DTI processing pipelines from the DWI datasets containing artefacts.
First, second, and third columns correspond to the 1st, 2nd, and 3rd DTI processing pipeline, respectively. First, second, third, and forth rows correspond to datasets LM-20, LM-40, SM-20, and SM-40, respectively. The corresponding T1-weighted structural image and full-brain segmentation of cerebral tissues are provided in panels e and f of Fig 2, respectively. The images were not interpolated.
Fig. 4. MD maps produced by the three proposed DTI processing pipelines from the DWI datasets containing artefacts.
MD maps produced by the three proposed DTI processing pipelines from the DWI datasets containing artefacts.
First, second, and third columns correspond to the 1st, 2nd, and 3rd DTI processing pipeline, respectively. First, second, third, and forth rows correspond to datasets LM-20, LM-40, SM-20, and SM-40, respectively. The corresponding T1-weighted structural image and full-brain segmentation of cerebral tissues are provided in panels e and f of Fig 2, respectively. The images were not interpolated.

Processing error of the proposed DTI pipelines

The differences in the measured FA and MD values associated with each pipeline when processing the DWI data without noise are provided in Fig 5. Considering FA and MD in both WM and GM regions, Pipeline #1 produced the most accurate FA and MD values.

Fig. 5. Processing error associated with the pipelines for FA calculation in (A) WM and in (B) GM, and MD calculation in (C) WM and in (D) GM.
Processing error associated with the pipelines for FA calculation in (A) WM and in (B) GM, and MD calculation in (C) WM and in (D) GM.

Histograms describing the voxelwise distribution of FA values in the WM region obtained using the three pipelines and the ground-truth data are provided in Fig 6. Despite a statistically significant difference between the pipelines (ANOVA, p<0.05) little difference is observed in the distribution of FA values.

Fig. 6. Distribution of FA values in the WM region generated by the three pipelines after processing the ground-truth data.
Distribution of FA values in the WM region generated by the three pipelines after processing the ground-truth data.

The Pearson correlation coefficient between the FA values in each voxel obtained with the 1st and 2nd pipelines, 1st and 3rd pipelines, and 2nd and 3rd pipelines were 0.939, 0.940, and 0.998, respectively. The scatter plots along with the corresponding least squares regression lines are provided in S1 to S3 Figs as supporting information.

Histograms describing the voxelwise distribution of the MD values in the WM region obtained using the three pipelines and the ground-truth data are provided in Fig 7. There was also a significant difference between the MD values obtained by the three pipelines (ANOVA, p<0.05).

Fig. 7. Distribution of the MD values in the WM region generated by the three pipelines after processing of the ground-truth data.
Distribution of the MD values in the WM region generated by the three pipelines after processing of the ground-truth data.

The Pearson correlation coefficient between the MD values in each voxel obtained with the 1st and 2nd pipelines, 1st and 3rd pipelines, and 2nd and 3rd pipelines were 0.978, 0.979, and 0.999, respectively. The scatter plots along with the corresponding least squares regression lines related to these correlation analyses are also provided in S4 to S6 Figs as supporting information.

Pipeline performance when processing DWI data containing artefacts

The differences in the measured FA and MD values in WM associated with each pipeline when processing the datasets containing artefacts is provided in Fig 8. In the case of FA, (Fig 8a), the percentage error did not differ significantly among the pipelines. In the case of MD, (Fig 8b), the percentage error associated with Pipeline #1 was lower than the other two pipelines.

Fig. 8. Percentage error associated with each pipeline when processing data containing artefacts.
Percentage error associated with each pipeline when processing data containing artefacts.
(A) shows FA in WM, (B) shows MD in WM region.

Performance of the pipelines in processing in vivo ONDRI DWI data

There was a significant difference in FA values measured by the pipelines in NAWM across the 20 ONDRI VCI subjects (p<0.01). Multiple comparison post hoc tests also showed that all pipelines differed significantly from each other (p<0.05) except pipeline-no QC and the 1st pipeline (p = 0.051), the 1st and 3rd pipelines (p = 0.051), and the 2nd and 3rd pipelines (p = 0.989). FA measurement variations in NAWM is also depicted in Fig 9.

Fig. 9. FA measurements in NAWM from 20 ONDRI VCI subjects.
FA measurements in NAWM from 20 ONDRI VCI subjects.

There was a significant difference in MD values measured by the pipelines in NAWM across the 20 ONDRI VCI subjects (p<0.01). Multiple comparison post hoc tests also showed that all pipelines did not differ significantly from each other (p>0.05), except the pipeline with no QC and the 1st pipeline (p<0.05) and the pipeline with no QC and the 3rd pipeline (p<0.05). It should be noted that the three proposed pipelines with diverse QC procedures (DTI-Prep and RESTORE) did not differ significantly with regard to MD measurement performance. MD measurements in NAWM made by all four pipelines are shown in Fig 10.

Fig. 10. MD measurements in NAWM from 20 ONDRI VCI subjects.
MD measurements in NAWM from 20 ONDRI VCI subjects.

Performance of the pipelines with an alternative denoising filter

The percentage error of the proposed pipelines for calculation of FA in WM when using the GAD filter in comparison to the LMMSE filter is depicted in Fig 11. For the purpose of qualitative comparisons the FA maps generated by the pipelines when using these two different filters were also presented in Fig 12. The use of the GAD filter increased the accuracy associated with all pipelines.

Fig. 11. Percentage error in calculation of FA in WM for the dataset LM-20 when using the gradient anisotropic diffusion (GAD) filter and Rician linear minimum mean square error (LMMSE) filter.
Percentage error in calculation of FA in WM for the dataset LM-20 when using the gradient anisotropic diffusion (GAD) filter and Rician linear minimum mean square error (LMMSE) filter.
All pipelines produced lower percentage error with the GAD filter.
Fig. 12. FA maps for the dataset LM-20 obtained by different pipelines when using the gradient anisotropic diffusion (GAD) filter (first row) and Rician linear minimum mean square error (LMMSE) filter (second row).
FA maps for the dataset LM-20 obtained by different pipelines when using the gradient anisotropic diffusion (GAD) filter (first row) and Rician linear minimum mean square error (LMMSE) filter (second row).
First, second, and third columns correspond to the 1st, 2nd, and 3rd DTI processing pipeline, respectively. The images were not interpolated.

Discussion

In this study, three different automated pipelines are presented for the analysis of the multisite ONDRI brain DTI data, each complying with the ENIGMA framework. The pipelines were designed by uniquely combining multiple image processing software tools to robustly and efficiently measure DTI metrics in the brain. In the proposed pipelines, three different QC procedures were utilized including robust tensor fitting (1st pipeline), the DTIPrep protocol (2nd pipeline), and the combination of both methods (3rd pipeline), to determine which produced optimal estimation of DTI metrics. It is noteworthy that based on the sequential application of these two QC procedures (DTIPrep and RESTORE) in the 3rd pipeline we expect that the overlap of the outliers detected by these algorithms would be minimal since these two algorithms affect the diffusion data on two different levels: the gradient volume level (in DTIPrep), and the voxel level (in RESTORE). These pipelines were applied to simulated DTI data to measure the processing error of the pipelines, to simulated DTI data containing artefacts to validate and compare the performance of the pipelines, and to 20 in vivo datasets from people with vascular cognitive impairment, to examine performance of the pipelines in a multi-site study.

All the pipelines successfully produced maps of the DTI metrics (FA, MD, AD, and RD maps) from the simulated ground-truth data and the simulated data containing noise and artifacts, the latter were qualitatively similar to the ground truth maps. The 1st pipeline produced the most accurate DTI metric values when compared to the ground truth, particularly for MD. Histograms showing the voxel wise distribution of FA and MD in NAWM obtained using the proposed pipelines (Figs 6 and 7) confirm that the performance of the three pipelines were similar for FA. Greater differences were observed between the 1st pipeline and the two other pipelines for MD.

The proposed pipelines also showed performance differences when tested in simulated datasets containing eddy current induced artifact, and different levels of motion artefacts and thermal noise. The performance of the pipelines was very similar for the calculation of FA in WM. However Pipeline #1 demonstrated the greatest accuracy in calculation of both FA and MD in WM (Fig 6). Both panels in Fig 8 also support the idea that the proposed pipelines successfully handle motion artifacts. In addition, noise had the expected effect on the calculation of DTI metrics. Specifically, with a higher level of noise (SNR = 20) there was a greater percentage error in DTI metrics calculations compared to lower noise levels (SNR = 40). Therefore we decided to test the effect of a denoising filter (GAD filter) in all the pipelines for the analysis of the artifactual dataset LM-20. As shown in Fig 11, all the pipelines presented considerably greater accuracy compared to the ground truth FA using the GAD filter in comparison with the Rician LMMSE filter used by default in the pipelines. This confirms that all the pipelines have the capacity to handle different levels of noise provided that an appropriate denoising filter is applied to the raw DWI data during processing.

The pipelines containing QC procedures were also used to analyze DWI datasets from 20 ONDRI VCI subjects and produced different results compared to the pipeline with no QC. The average FA value in the NAWM region measured in the current study was 0.39 ± 0.03. This value is in agreement with previously published measurements of average FA in WM regions and specific WM tracts in VCI subjects [8385]. Specifically, the average FA value in the WM of VCI subjects was previously reported as 0.41 ± 0.026 [83]. All three pipelines performed similarly with regard to MD measurements while only the 2nd and 3rd pipelines performed similarly with respect to FA measurements. The 3rd pipeline was the most comprehensive as it included a combination of two different QC procedures and produced the highest precision (i.e., lower STD) for FA calculation in NAWM (Fig 9). High measurement precision is an important feature for the analysis of human cohort data where the intention is to detect pathological microstructural alterations among different neurodegenerative conditions towards more effective diagnosis and treatment of these diseases.

Study limitations

In the current study, we registered the DTI data to the high resolution structural image before tensor fitting, which is consistent with the well-known ENIGMA protocol [23,24,86]. The primary reason was that reverse phase encode direction DTI data was not available as part of the ONDRI DTI data acquisition, and hence registering to the native T1 space provided a major benefit to diminish EPI-related artefacts (distortions). The pipelines in the current manuscript were developed for the calculation of scalar DTI metrics. These metrics are expected to be minimally affected by issues arising from signal misorientation, which are more relevant for tractography applications. The current study utilized the BBR approach to directly register DTI to the native T1-weighted image. An alternative approach is to first register the DTI data to the corresponding T2-weighted image using the BBR method followed by a linear registration to the T1-weighted image as previously described [87]. However, this additional linear registration would further interpolate the DTI data, which may not be preferred but could be considered as an alternative approach if the direct registration to the native T1 space was unsatisfactory.

Conclusions

Three different automated DTI processing pipelines, complying with the ENIGMA framework were fully implemented and then evaluated using simulated ground truth and artifact containing DWI datasets as well as multisite DTI data from ONDRI. These unique pipelines were created by optimizing and connecting components from seven different image processing software packages. The major difference between the proposed pipelines was in the approach to quality control. All the pipelines produced similar FA measurements while the 1st pipeline incorporating the RESTORE algorithm produced more accurate MD measurements. The performance of the 2nd and 3rd pipelines was very similar for both FA and MD measurements. Based on the performance of the pipelines in the analysis of the DTI data from the ONDRI database and the observed DTI metric measurement precision, we conclude that the pipeline that combined both the DTIPrep and RESTORE algorithms is more comprehensive and hence recommended for the analysis of multisite DTI data.

Supporting information

S1 Table [docx]
ONDRI Subject IDs and Scan Dates related to the vascular cognitive impairment (VCI) subjects used in this study.

S2 Table [docx]
Percentage error in FA produced by each pipeline when processing the ground-truth data and the simulated datasets with artifacts.

S3 Table [docx]
Percentage error in MD produced by each pipeline when processing the ground-truth data and the simulated datasets with artifacts.

S1 Fig [docx]
Scatter plot along with least squares fitted line representing correlation between WM FA measurement in the 1 and 2 pipelines when they were utilized for processing ground-truth data (β = 0.939).

S2 Fig [docx]
Scatter plot along with least squares fitted line representing correlation between WM FA measurement in the 1 and 3 pipelines when they were utilized for processing ground-truth data (β = 0.940).

S3 Fig [docx]
Scatter plot along with least squares fitted line representing correlation between WM FA measurement in the 2 and 3 pipelines when they were utilized for processing ground-truth data (β = 0.998).

S4 Fig [docx]
Scatter plot along with least squares fitted line representing correlation between WM MD measurement in the 1 and 2 pipelines when they were utilized for processing ground-truth data (β = 0.978).

S5 Fig [docx]
Scatter plot along with least squares fitted line representing correlation between WM MD measurement in the 1 and 3 pipelines when they were utilized for processing ground-truth data (β = 0.979).

S6 Fig [docx]
Scatter plot along with least squares fitted line representing correlation between WM MD measurement in the 2 and 3 pipelines when they were utilized for processing ground-truth data (β = 0.999).


Zdroje

1. Alexander AL, Lee JE, Lazar M, Field AS. Diffusion tensor imaging of the brain. Neurotherapeutics. 2007;4: 316–29. doi: 10.1016/j.nurt.2007.05.011 17599699

2. Rajagopalan V, Jiang Z, Yue GH, Radic JS, Pioro EP, Wylie GR, et al. A Basic Introduction to Diffusion Tensor Imaging Mathematics and Image Processing Steps. Brain Disord Ther. 2017;6.

3. Hulkower MB, Poliak DB, Rosenbaum SB, Zimmerman ME, Lipton ML. A Decade of DTI in Traumatic Brain Injury: 10 Years and 100 Articles Later. Am J Neuroradiol. 2013;34: 2064–2074. doi: 10.3174/ajnr.A3395 23306011

4. Beaulieu C. The basis of anisotropic water diffusion in the nervous system—a technical review. NMR Biomed. 2002;15: 435–455. doi: 10.1002/nbm.782 12489094

5. Alfaro-Almagro F, Jenkinson M, Bangerter NK, Andersson JLR, Griffanti L, Douaud G, et al. Image processing and Quality Control for the first 10,000 brain imaging datasets from UK Biobank. Neuroimage. 2018;166: 400–424. doi: 10.1016/j.neuroimage.2017.10.034 29079522

6. Budin F, Hoogstoel M, Reynolds P, Grauer M, O’Leary-Moore SK, Oguz I. Fully automated rodent brain MR image processing pipeline on a Midas server: from acquired images to region-based statistics. Front Neuroinform. 2013;7: 15. doi: 10.3389/fninf.2013.00015 23964234

7. Cui Z, Zhong S, Xu P, He Y, Gong G. PANDA: a pipeline toolbox for analyzing brain diffusion images. Front Hum Neurosci. 2013;7: 42. doi: 10.3389/fnhum.2013.00042 23439846

8. Smith SM, Jenkinson M, Johansen-Berg H, Rueckert D, Nichols TE, Mackay CE, et al. Tract-based spatial statistics: Voxelwise analysis of multi-subject diffusion data. Neuroimage. 2006;31: 1487–1505. doi: 10.1016/j.neuroimage.2006.02.024 16624579

9. Jahanshad N, Kochunov P V., Sprooten E, Mandl RC, Nichols TE, Almasy L, et al. Multi-site genetic analysis of diffusion images and voxelwise heritability analysis: A pilot project of the ENIGMA–DTI working group. Neuroimage. 2013;81: 455–469. doi: 10.1016/j.neuroimage.2013.04.061 23629049

10. Glasser MF, Sotiropoulos SN, Wilson JA, Coalson TS, Fischl B, Andersson JL, et al. The minimal preprocessing pipelines for the Human Connectome Project. Neuroimage. 2013;80: 105–124. doi: 10.1016/j.neuroimage.2013.04.127 23668970

11. Jones DK, Cercignani M. Twenty-five pitfalls in the analysis of diffusion MRI data. NMR Biomed. 2010;23: 803–820. doi: 10.1002/nbm.1543 20886566

12. Jiang H, van Zijl PCM, Kim J, Pearlson GD, Mori S. DtiStudio: Resource program for diffusion tensor computation and fiber bundle tracking. Comput Methods Programs Biomed. 2006;81: 106–116. doi: 10.1016/j.cmpb.2005.08.004 16413083

13. Oguz I, Farzinfar M, Matsui J, Budin F, Liu Z, Gerig G, et al. DTIPrep: quality control of diffusion-weighted images. Front Neuroinform. 2014;8: 4. doi: 10.3389/fninf.2014.00004 24523693

14. Koay CG, Özarslan E, Pierpaoli C. Probabilistic Identification and Estimation of Noise (PIESNO): A self-consistent approach and its applications in MRI. J Magn Reson. 2009;199: 94–103. doi: 10.1016/j.jmr.2009.03.005 19346143

15. Mirzaalian H, Ning L, Savadjiev P, Pasternak O, Bouix S, Michailovich O, et al. Inter-site and inter-scanner diffusion MRI data harmonization. Neuroimage. 2016;135: 311–323. doi: 10.1016/j.neuroimage.2016.04.041 27138209

16. Fortin J-P, Parker D, Tunç B, Watanabe T, Elliott MA, Ruparel K, et al. Harmonization of multi-site diffusion tensor imaging data. Neuroimage. 2017;161: 149–170. doi: 10.1016/j.neuroimage.2017.08.047 28826946

17. Jenkins J, Chang L-C, Hutchinson E, Irfanoglu MO, Pierpaoli C. Harmonization of methods to facilitate reproducibility in medical data processing: Applications to diffusion tensor magnetic resonance imaging. 2016 IEEE International Conference on Big Data (Big Data). IEEE; 2016. pp. 3992–3994.

18. Suinesiaputra A, Sanghvi MM, Aung N, Paiva JM, Zemrak F, Fung K, et al. Fully-automated left ventricular mass and volume MRI analysis in the UK Biobank population cohort: evaluation of initial results. Int J Cardiovasc Imaging. 2017 [cited 1 Jan 2018]. doi: 10.1007/s10554-017-1225-9 28836039

19. Welikala RA, Foster PJ, Whincup PH, Rudnicka AR, Owen CG, Strachan DP, et al. Automated arteriole and venule classification using deep learning for retinal images from the UK Biobank cohort. Comput Biol Med. 2017;90: 23–32. doi: 10.1016/j.compbiomed.2017.09.005 28917120

20. Petersen SE, Sanghvi MM, Aung N, Cooper JA, Paiva JM, Zemrak F, et al. The impact of cardiovascular risk factors on cardiac structure and function: Insights from the UK Biobank imaging enhancement study. Fukumoto Y, editor. PLoS One. 2017;12: e0185114. doi: 10.1371/journal.pone.0185114 28973022

21. Farhan SMK, Bartha R, Black SE, Corbett D, Finger E, Freedman M, et al. The Ontario Neurodegenerative Disease Research Initiative (ONDRI). Can J Neurol Sci / J Can des Sci Neurol. 2017;44: 196–202. doi: 10.1017/cjn.2016.415 28003035

22. Lauzon CB, Asman AJ, Esparza ML, Burns SS, Fan Q, Gao Y, et al. Simultaneous Analysis and Quality Assurance for Diffusion Tensor Imaging. Alexander DC, editor. PLoS One. 2013;8: e61737. doi: 10.1371/journal.pone.0061737 23637895

23. Thompson PM, Stein JL, Medland SE, Hibar DP, Vasquez AA, Renteria ME, et al. The ENIGMA Consortium: large-scale collaborative analyses of neuroimaging and genetic data. Brain Imaging Behav. 2014;8: 153–82. doi: 10.1007/s11682-013-9269-5 24399358

24. Kochunov P, Jahanshad N, Sprooten E, Nichols TE, Mandl RC, Almasy L, et al. Multi-site study of additive genetic effects on fractional anisotropy of cerebral white matter: Comparing meta and megaanalytical approaches for data pooling. Neuroimage. 2014;95: 136–150. doi: 10.1016/j.neuroimage.2014.03.033 24657781

25. Van Essen DC, Ugurbil K, Auerbach E, Barch D, Behrens TEJ, Bucholz R, et al. The Human Connectome Project: A data acquisition perspective. Neuroimage. 2012;62: 2222–2231. doi: 10.1016/j.neuroimage.2012.02.018 22366334

26. Farzinfar M, Oguz I, Smith RG, Verde AR, Dietrich C, Gupta A, et al. Diffusion imaging quality control via entropy of principal direction distribution. Neuroimage. 2013;82: 1–12. doi: 10.1016/j.neuroimage.2013.05.022 23684874

27. Chang L-C, Jones DK, Pierpaoli C. RESTORE: Robust estimation of tensors by outlier rejection. Magn Reson Med. 2005;53: 1088–1095. doi: 10.1002/mrm.20426 15844157

28. Graham MS, Drobnjak I, Zhang H. Realistic simulation of artefacts in diffusion MRI for validating post-processing correction techniques. Neuroimage. 2016;125: 1079–1094. doi: 10.1016/j.neuroimage.2015.11.006 26549300

29. Graham MS, Drobnjak I, Jenkinson M, Zhang H. Quantitative assessment of the susceptibility artefact and its interaction with motion in diffusion MRI. Yap P-T, editor. PLoS One. 2017;12: e0185647. doi: 10.1371/journal.pone.0185647 28968429

30. Drobnjak I, Gavaghan D, Süli E, Pitt-Francis J, Jenkinson M. Development of a functional magnetic resonance imaging simulator for modeling realistic rigid-body motion artifacts. Magn Reson Med. 2006;56: 364–380. doi: 10.1002/mrm.20939 16841304

31. Drobnjak I, Pell GS, Jenkinson M. Simulating the effects of time-varying magnetic fields with a realistic simulated scanner. Magn Reson Imaging. 2010;28: 1014–1021. doi: 10.1016/j.mri.2010.03.029 20418038

32. NITRC: Simulated Diffusion-Weighted Datasets: Tool/Resource Info. [cited 10 Apr 2019]. https://www.nitrc.org/projects/diffusionsim/

33. Zhang Y, Brady M, Smith S. Segmentation of brain MR images through a hidden Markov random field model and the expectation-maximization algorithm. IEEE Trans Med Imaging. 2001;20: 45–57. doi: 10.1109/42.906424 11293691

34. Alexander DC, Barker GJ, Arridge SR. Detection and modeling of non-Gaussian apparent diffusion coefficient profiles in human brain data. Magn Reson Med. 2002;48: 331–340. doi: 10.1002/mrm.10209 12210942

35. Montero-Odasso M, Pieruccini-Faria F, Bartha R, Black SE, Finger E, Freedman M, et al. Motor Phenotype in Neurodegenerative Disorders: Gait and Balance Platform Study Design Protocol for the Ontario Neurodegenerative Research Initiative (ONDRI). J Alzheimer’s Dis. 2017;59: 707–721. doi: 10.3233/JAD-170149 28671116

36. Dade L., Gao F., Kovacevic N, Roy P, Rockel C, O’Toole C., et al. Semiautomatic brain region extraction: a method of parcellating brain regions from structural magnetic resonance images. Neuroimage. 2004;22: 1492–1502. doi: 10.1016/j.neuroimage.2004.03.023 15275906

37. Ramirez J, Scott CJM, McNeely AA, Berezuk C, Gao F, Szilagyi GM, et al. Lesion Explorer: a video-guided, standardized protocol for accurate and reliable MRI-derived volumetrics in Alzheimer’s disease and normal elderly. J Vis Exp. 2014 [cited 10 Jul 2018]. doi: 10.3791/50887 24797507

38. Ramirez J, Gibson E, Quddus A, Lobaugh NJ, Feinstein A, Levine B, et al. Lesion Explorer: A comprehensive segmentation and parcellation package to obtain regional volumetrics for subcortical hyperintensities and intracranial tissue. Neuroimage. 2011;54: 963–973. doi: 10.1016/j.neuroimage.2010.09.013 20849961

39. Gibson E, Gao F, Black SE, Lobaugh NJ. Automatic segmentation of white matter hyperintensities in the elderly using FLAIR images at 3T. J Magn Reson Imaging. 2010;31: 1311–1322. doi: 10.1002/jmri.22004 20512882

40. Ramirez J, Scott CJM, Black SE. A Short-Term Scan–Rescan Reliability Test Measuring Brain Tissue and Subcortical Hyperintensity Volumetrics Obtained Using the Lesion Explorer Structural MRI Processing Pipeline. Brain Topogr. 2013;26: 35–38. doi: 10.1007/s10548-012-0228-z 22562092

41. Kovacevic N, Lobaugh NJ, Bronskill MJ, Levine B, Feinstein A, Black SE. A robust method for extraction and automatic segmentation of brain images. Neuroimage. 2002;17: 1087–100. Available: http://www.ncbi.nlm.nih.gov/pubmed/12414252 doi: 10.1006/nimg.2002.1221

42. Ramirez J, Berezuk C, McNeely AA, Scott CJM, Gao F, Black SE. Visible Virchow-Robin Spaces on Magnetic Resonance Imaging of Alzheimer’s Disease Patients and Normal Elderly from the Sunnybrook Dementia Study. J Alzheimer’s Dis. 2014;43: 415–424. doi: 10.3233/JAD-132528 25096616

43. Fedorov A, Beichel R, Kalpathy-Cramer J, Finet J, Fillion-Robin J-C, Pujol S, et al. 3D Slicer as an image computing platform for the Quantitative Imaging Network. Magn Reson Imaging. 2012;30: 1323–1341. doi: 10.1016/j.mri.2012.05.001 22770690

44. Jenkinson M, Beckmann CF, Behrens TEJ, Woolrich MW, Smith SM. FSL. Neuroimage. 2012;62: 782–790. doi: 10.1016/j.neuroimage.2011.09.015 21979382

45. Avants BB, Tustison NJ, Song G, Cook PA, Klein A, Gee JC. A reproducible evaluation of ANTs similarity metric performance in brain image registration. Neuroimage. 2011;54: 2033–44. doi: 10.1016/j.neuroimage.2010.09.025 20851191

46. Cook PA, Bai Y, Nedjati-Gilani S, Seunarine KK, Hall MD, Parker GJM, et al. Camino: Open-Source Diffusion-MRI Reconstruction and Processing. 2006. https://www.semanticscholar.org/paper/Camino%3A-Open-Source-Diffusion-MRI-Reconstruction-Cook-Bai/6de9805ed125e5479e4c58f82472aac2ab6d85df

47. Rorden C, Brett M. Stereotaxic Display of Brain Lesions. Behav Neurol. 2000;12: 191–200. doi: 10.1155/2000/421719 11568431

48. Adluru N, Zhang H, Fox AS, Shelton SE, Ennis CM, Bartosic AM, et al. A diffusion tensor brain template for Rhesus Macaques. Neuroimage. 2012;59: 306–318. doi: 10.1016/j.neuroimage.2011.07.029 21803162

49. Mustra M, Delac K, Grgic M. Overview of the DICOM standard. 2008 50th International Symposium ELMAR. 2008. pp. 39–44.

50. Cox RW, Ashburner J, Breman H, Fissell K, Haselgrove C, Holmes CJ, et al. A (sort of) new image data format standard: NiFTI-1. 10th Annual Meeting of the Organization for Human Brain Mapping. 2004.

51. 3D Slicer. [cited 17 Apr 2018]. https://www.slicer.org/

52. Kikinis R, Pieper SD, Vosburgh KG. 3D Slicer: A Platform for Subject-Specific Image Analysis, Visualization, and Clinical Support. Intraoperative Imaging and Image-Guided Therapy. New York, NY: Springer New York; 2014. pp. 277–289. doi: 10.1007/978-1-4614-7657-3_19

53. MRIcron Index Page. [cited 17 Apr 2018]. http://people.cas.sc.edu/rorden/mricron/index.html

54. Jovicich J, Czanner S, Greve D, Haley E, van der Kouwe A, Gollub R, et al. Reliability in multi-site structural MRI studies: Effects of gradient non-linearity correction on phantom and human data. Neuroimage. 2006;30: 436–443. doi: 10.1016/j.neuroimage.2005.09.046 16300968

55. Collewet G, Davenel A, Toussaint C, Akoka S. Correction of intensity nonuniformity in spin-echo T(1)-weighted images. Magn Reson Imaging. 2002;20: 365–73. doi: 10.1016/s0730-725x(02)00502-7 12165356

56. Tustison NJ, Avants BB, Cook PA, Yuanjie Zheng Y, Egan A, Yushkevich PA, et al. N4ITK: Improved N3 Bias Correction. IEEE Trans Med Imaging. 2010;29: 1310–1320. doi: 10.1109/TMI.2010.2046908 20378467

57. Iglesias JE, Liu Cheng-Yi, Thompson PM, Tu Zhuowen. Robust Brain Extraction Across Datasets and Comparison With Publicly Available Methods. IEEE Trans Med Imaging. 2011;30: 1617–1634. doi: 10.1109/TMI.2011.2138152 21880566

58. Huang H, Ceritoglu C, Li X, Qiu A, Miller MI, van Zijl PCM, et al. Correction of B0 susceptibility induced distortion in diffusion-weighted images using large-deformation diffeomorphic metric mapping. Magn Reson Imaging. 2008;26: 1294–1302. doi: 10.1016/j.mri.2008.03.005 18499384

59. Bhushan C, Haldar JP, Joshi AA, Leahy RM. Correcting Susceptibility-Induced Distortion in Diffusion-Weighted MRI using Constrained Nonrigid Registration. Signal Inf Process Assoc Annu Summit Conf (APSIPA). Asia-Pacific Asia-Pacific Signal Inf Process Assoc Annu Summit Conf. 2012;2012. http://www.ncbi.nlm.nih.gov/pubmed/26767197

60. Kybic J, Thevenaz P, Nirkko A, Unser M. Unwarping of unidirectionally distorted EPI images. IEEE Trans Med Imaging. 2000;19: 80–93. doi: 10.1109/42.836368 10784280

61. Jenkinson M, Bannister P, Brady M, Smith S. Improved optimization for the robust and accurate linear registration and motion correction of brain images. Neuroimage. 2002;17: 825–41. http://www.ncbi.nlm.nih.gov/pubmed/12377157 doi: 10.1016/s1053-8119(02)91132-8

62. Jenkinson M, Smith S. A global optimisation method for robust affine registration of brain images. Med Image Anal. 2001;5: 143–56. Available: http://www.ncbi.nlm.nih.gov/pubmed/11516708 doi: 10.1016/s1361-8415(01)00036-6

63. Greve DN, Fischl B. Accurate and robust brain image alignment using boundary-based registration. Neuroimage. 2009;48: 63–72. doi: 10.1016/j.neuroimage.2009.06.060 19573611

64. Basu S, Fletcher T, Whitaker R. Rician noise removal in diffusion tensor MRI. Med Image Comput Comput Assist Interv. 2006;9: 117–25. Available: http://www.ncbi.nlm.nih.gov/pubmed/17354881

65. Tristán-Vega A, Aja-Fernández S. DWI filtering using joint information for DTI and HARDI. Med Image Anal. 2010;14: 205–218. doi: 10.1016/j.media.2009.11.001 20005152

66. Aja-Fernandez S, Niethammer M, Kubicki M, Shenton ME, Westin C-F. Restoration of DWI data using a Rician LMMSE estimator. IEEE Trans Med Imaging. 2008;27: 1389–403. doi: 10.1109/TMI.2008.920609 18815091

67. Polzehl J, Tabelow K. Structural adaptive smoothing Diffusion Tensor Imaging data: the R Package dti. Neuroimage. 2009;47: S51. doi: 10.1016/S1053-8119(09)70129-6

68. Andersson JLR, Skare S, Ashburner J. How to correct susceptibility distortions in spin-echo echo-planar images: application to diffusion tensor imaging. Neuroimage. 2003;20: 870–888. doi: 10.1016/S1053-8119(03)00336-7 14568458

69. Andersson JLR, Sotiropoulos SN. An integrated approach to correction for off-resonance effects and subject movement in diffusion MR imaging. Neuroimage. 2016;125: 1063–1078. doi: 10.1016/j.neuroimage.2015.10.019 26481672

70. Gallichan D, Andersson JLR, Jenkinson M, Robson MD, Miller KL. Reducing distortions in diffusion-weighted echo planar imaging with a dual-echo blip-reversed sequence. Magn Reson Med. 2010;64: n/a–n/a. doi: 10.1002/mrm.22318 20665782

71. Wang S, Peterson DJ, Gatenby JC, Li W, Grabowski TJ, Madhyastha TM, et al. Evaluation of Field Map and Nonlinear Registration Methods for Correction of Susceptibility Artifacts in Diffusion MRI. 2017 [cited 17 Apr 2018].

72. Calhoun VD, Wager TD, Krishnan A, Rosch KS, Seymour KE, Nebel MB, et al. The Impact of T1 Versus EPI Spatial Normalization Templates for fMRI Data Analyses. Hum Brain Mapp. 2017;38: 5331–5342. doi: 10.1002/hbm.23737 28745021

73. Klein A, Andersson J, Ardekani BA, Ashburner J, Avants B, Chiang M-C, et al. Evaluation of 14 nonlinear deformation algorithms applied to human brain MRI registration. Neuroimage. 2009;46: 786–802. doi: 10.1016/j.neuroimage.2008.12.037 19195496

74. Avants BB, Epstein CL, Grossman M, Gee JC. Symmetric diffeomorphic image registration with cross-correlation: Evaluating automated labeling of elderly and neurodegenerative brain. Med Image Anal. 2008;12: 26–41. doi: 10.1016/j.media.2007.06.004 17659998

75. Smith SM. Fast robust automated brain extraction. Hum Brain Mapp. 2002;17: 143–155. doi: 10.1002/hbm.10062 12391568

76. Camino | UCL Camino Diffusion MRI Toolkit. [cited 20 Apr 2018]. http://camino.cs.ucl.ac.uk/index.php?n=Main.HomePage

77. Basser PJ, Mattiello J, Lebihan D. Estimation of the Effective Self-Diffusion Tensor from the NMR Spin Echo. J Magn Reson Ser B. 1994;103: 247–254. doi: 10.1006/JMRB.1994.1037 8019776

78. Behrens TEJ, Woolrich MW, Jenkinson M, Johansen-Berg H, Nunes RG, Clare S, et al. Characterization and propagation of uncertainty in diffusion-weighted MR imaging. Magn Reson Med. 2003;50: 1077–1088. doi: 10.1002/mrm.10609 14587019

79. Behrens TEJ, Berg HJ, Jbabdi S, Rushworth MFS, Woolrich MW. Probabilistic diffusion tractography with multiple fibre orientations: What can we gain? Neuroimage. 2007;34: 144–155. doi: 10.1016/j.neuroimage.2006.09.018 17070705

80. Moonen XJEF, Foster-Dingley XJC, Van Den Berg-Huijsmans XAA, De Ruijter XW, De Craen XAJM, Van Der Grond XJ, et al. Influence of Small Vessel Disease and Microstructural Integrity on Neurocognitive Functioning in Older Individuals: The DANTE Study Leiden. [cited 20 Apr 2018]. doi: 10.3174/ajnr.A4934 27659190

81. Weickert J, Stuttgart BGT. Anisotropic Diffusion in Image Processing. https://www.mia.uni-saarland.de/weickert/Papers/book.pdf

82. Palma CA, Cappabianco FAM, Ide JS, Miranda PAV. Anisotropic Diffusion Filtering Operation and Limitations—Magnetic Resonance Imaging Evaluation. IFAC Proc Vol. 2014;47: 3887–3892. doi: 10.3182/20140824-6-ZA-1003.02347

83. Correia S, Lee SY, Voorn T, Tate DF, Paul RH, Zhang S, et al. Quantitative Tractography Metrics of White Matter Integrity in Diffusion-Tensor MRI. [cited 23 Jul 2019]. doi: 10.1016/j.neuroimage.2008.05.022 18617421

84. Chen H-J, Gao Y-Q, Che C-H, Lin H, Ruan X-L. Diffusion Tensor Imaging With Tract-Based Spatial Statistics Reveals White Matter Abnormalities in Patients With Vascular Cognitive Impairment. Front Neuroanat. 2018;12: 53. doi: 10.3389/fnana.2018.00053 29997482

85. Biesbroek JM, Leemans A, den Bakker H, Duering M, Gesierich B, Koek HL, et al. Microstructure of Strategic White Matter Tracts and Cognition in Memory Clinic Patients with Vascular Brain Injury. Dement Geriatr Cogn Disord. 2017;44: 268–282. doi: 10.1159/000485376 29353280

86. DTI Protocols « ENIGMA. [cited 1 Oct 2019]. http://enigma.ini.usc.edu/protocols/dti-protocols/

87. Bastiani M, Andersson JLR, Cordero-Grande L, Murgasova M, Hutter J, Price AN, et al. Automated processing pipeline for neonatal diffusion MRI in the developing Human Connectome Project. Neuroimage. 2019;185: 750–763. doi: 10.1016/j.neuroimage.2018.05.064 29852283


Článok vyšiel v časopise

PLOS One


2019 Číslo 12
Najčítanejšie tento týždeň
Najčítanejšie v tomto čísle
Kurzy

Zvýšte si kvalifikáciu online z pohodlia domova

Aktuální možnosti diagnostiky a léčby litiáz
nový kurz
Autori: MUDr. Tomáš Ürge, PhD.

Všetky kurzy
Prihlásenie
Zabudnuté heslo

Zadajte e-mailovú adresu, s ktorou ste vytvárali účet. Budú Vám na ňu zasielané informácie k nastaveniu nového hesla.

Prihlásenie

Nemáte účet?  Registrujte sa

#ADS_BOTTOM_SCRIPTS#