Systematic review of the accuracy of plasma preparation tubes for HIV viral load testing


Authors: Robert Luo aff001;  Jessica Markby aff002;  Jilian Sacks aff002;  Lara Vojnov aff001
Authors place of work: World Health Organization, Geneva, Switzerland aff001;  Clinton Health Access Initiative, Boston, Massachusetts, United States of America aff002
Published in the journal: PLoS ONE 14(11)
Category: Research Article
doi: https://doi.org/10.1371/journal.pone.0225393

Summary

Expanding access to HIV viral load testing is essential to improving the care and treatment of people living with HIV/AIDS and ending the AIDS epidemic. Though significant investments have been made in the past five years, many high burden, low resource countries continue to have viral load access rates below 50%. Plasma preparation tubes (PPTs) can simplify storage, transport, and preparation of plasma used for viral load testing. A systematic review was conducted to evaluate the accuracy of plasma preparation tubes for HIV viral load testing. Study results regarding the accuracy of PPT viral load measurements across various storage and transportation conditions were examined. The quality of evidence was evaluated using GRADE and QUADAS-2 criteria. The review identified 16 studies using PPTs with data from 6,141 individuals from 1995 to 2014. Overall the quality of evidence was rated as moderate, with unclear applicability for studies evaluating viral load assays that are no longer commercially available. Significantly elevated viral load results (>0.3 log copies/ml difference) have been observed with PPTs; however, when manufacturer handling instructions are followed, when plasma is aliquoted into a secondary tube, or when PPTs are centrifuged prior to testing, PPT results only differed from standard EDTA plasma testing using commercially available viral load assays by a range on average of -0.03 to +0.08 log copies/ml across studies. Although spuriously elevated viral load results have been observed with PPTs, following proper sample handing techniques have been shown to provide accurate results. PPTs, therefore, provide a high quality alternative specimen type for countries seeking solutions to infrastructure and specimen transportation challenges in an effort to scale-up viral load testing and achieve 90-90-90 targets.

Keywords:

Blood – blood plasma – Viral load – HIV-1 – Systematic reviews – United States – Centrifugation

Introduction

Viral load testing is the preferred method for monitoring individuals on HIV antiretroviral treatment [1]. Expanding access to viral load is a priority for improving the care and treatment of persons with HIV/AIDS, particularly those living in resource limited settings, and is essential for meeting the UNAIDS 90-90-90 goals for ending the AIDS epidemic [2]. The primary and preferred sample type for viral load testing is plasma [3], which is prepared from whole blood collected via venipuncture. Often whole blood is collected into an EDTA (ethylenediaminetetraacetic acid anti-coagulant) tube (purple or lavender cap). Whole blood in EDTA tubes, however, must be transported and centrifuged into plasma within 6–24 hours of sample collection, depending on the manufacturer [4]. This can be restrictive for many countries and health care facilities. Though investments and viral load scale-up have been significant since 2013, access has been limited in high burden countries to those with advanced infrastructure and robust specimen transport networks [5,6]. Dried blood spots, which can be prepared from fingerstick whole blood, provide an alternative method for sample collection; however, limited progress has been made in ensuring the quality of using dried blood spots for HIV viral load testing through international regulatory approvals [7]. Because of this, countries have sought alternative specimen types and/or technologies to further support and expand viral load testing access.

Unlike standard EDTA blood collection tubes, plasma preparation tubes (PPTs) can facilitate simpler handling and storage of plasma for HIV viral load testing. PPTs use the same EDTA anticoagulant, but contain a gel that separates the plasma from blood cells during centrifugation [8]. After the blood is collected, the PPT is spun in a centrifuge within 24 hours and a gel barrier inside the PPT separates the plasma from the rest of the whole blood so the plasma can be used for HIV viral load testing. The same plasma specimen volume is used for the viral load assay; therefore, limits of detection are synonymous with EDTA plasma. Additionally, plasma can be stored in the PPTs in situ if desired, without the need for additional manual pipetting steps to a secondary tube, which can reduce the risk of sample handling errors and the potential for cross-contamination [9].

A recent systematic review on blood stability indicated that PPTs are likely stable for at least 5–7 days, far beyond manufacturer instructions, depending on storage temperatures [9]. However, despite their growing and widespread use in many countries, including high burden, low resource settings such as Kenya and South Africa, and their ability to increase access to viral load testing, there has not yet been a systematic review examining the accuracy of PPTs for HIV viral load testing. Additionally, conflicting results reported in the literature with regards to their accuracy and optimal handling methods have raised questions concerning their use with different manufacturers’ viral load assays. Therefore, a systematic review was conducted to evaluate the accuracy of PPTs for HIV viral load testing.

Methods

A systematic review was performed according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA, completed checklist in S1 Table) [10] following a pre-defined study protocol. Three databases [PubMed/Medline, International AIDS Society conference proceedings, and CROI (Conference on Retroviruses and Opportunistic Infections) conference proceedings] were searched using search terms from the protocol (detailed in the S1 File) in January 2019 to identify potentially relevant studies. No restrictions on publication year, publication status, or language were used.

Two reviewers independently screened all titles and abstracts for inclusion and reviewed all potentially relevant studies in full. Studies were included if they evaluated the accuracy of PPTs for HIV viral load testing by comparing viral load results from PPTs to viral load results from either EDTA tubes or PPT tubes tested on another viral load assay approved by a stringent regulatory authority. Studies were excluded if they did not include a proper comparator or if they were purely review articles.

Data were extracted and summarized for all included studies, outlining the study design, methods, and principle components of each study (e.g., sample size, viral load assay used, storage and transport conditions of samples). The primary outcome assessed was accuracy of the PPT measurements, defined by studies as a significant statistical or clinical difference between the PPT viral load measurements and the comparator viral load result used in the study. The quality and risk of bias of individual studies were assessed using QUADAS-2 criteria [11], and the entire body of evidence included in the review was assessed using GRADE criteria [12].

Results

The review identified forty-two studies through database searches, of which sixteen met the inclusion criteria (Fig 1). Overall the studies were published between 1995–2014 and included samples from 6,141 individuals. Although the majority of studies (11/16, 69%) were conducted in the United States, the majority of samples (4049/6141, 66%) included came from a large retrospective study done in Norway. Due to the vast difference in sample storage and transport conditions among studies, the results were only qualitatively summarized, and a meta-analysis was not undertaken. The main characteristics for each study are summarized in Table 1.

Fig. 1. PRISMA diagram of studies.
PRISMA diagram of studies.
Tab. 1. Summary of Studies.
Summary of Studies.

The quality of studies included in the systematic review was moderate overall. Evaluation of each study following QUADAS-2 criteria showed a low risk of bias. Samples were taken from patients with a wide range of HIV viral loads. Of note, the majority of studies (14/16, 88%) were either done in North America or Western Europe, with only two of the studies including samples from high burden HIV regions. However, that did not affect the risk of bias, as studies evaluated a broad range of sample handling conditions, including transportation from remote clinics. A majority of studies (12/16, 75%) also had unclear applicability for the index and reference tests used, since the tests were no longer commercially available and have been replaced by newer tests (detailed in S2 Table). GRADE assessment did not reveal any serious risk of bias, indirectness or imprecision, although inconsistency was high across studies due to the variability in storage and transportation conditions (Table 2).

Tab. 2. GRADE evaluation of evidence quality.
GRADE evaluation of evidence quality.

The earliest group of studies were published between 1995 and 2005 and validated that PPTs could be used with no significant differences in viral load results [1316]. However, later studies demonstrated elevated viral loads using PPTs (over a 0.3 log copies/ml increase compared to standard EDTA plasma), particularly at viral loads less than 5,000 copies/ml [1719]. The elevated viral load results were likely due to the leakage of cells containing HIV nucleic acids, such as proviral HIV DNA and intracellular RNA, from the cellular component of whole blood, that moved back through the gel barrier into the plasma. Additional studies found this issue could be resolved by either aliquoting the plasma into a secondary tube after the initial centrifugation [2022], or repeating centrifugation after transport of the PPTs to the lab prior to aliquoting and testing [2324].

Four published studies evaluated PPTs on viral load assays that were commercially available at the time of this review (Abbott RealTime HIV-1 and Roche COBAS AmpliPREP/COBAS TaqMan HIV-1 Test, v2.0) [2528]. The three studies evaluating the Abbott viral load assay showed no significant change in viral load results regardless of if the PPTs were frozen and thawed or transported after initial centrifugation and prior to testing [2527]. The three studies assessing the Roche assay found elevated viral load results if the PPTs were frozen or transported without a second centrifugation prior to testing [2628]. These viral load results were found to be up to 5,260 copies/ml higher than plasma prepared from a standard EDTA collection tube, with the differences being most noticeable for plasma viral loads less than 1,000 copies/ml due to their relative magnitude compared to the standard EDTA viral load result. Consequently, manufacturer instructions recommend an additional centrifugation step prior to testing using the Roche assay. For both the Abbott and Roche assays, aliquoting the plasma into a secondary tube after initial centrifugation also ensured accurate viral load results. Table 3 summarizes the methods published in the literature that support accurate results for the current use of PPTs. When proper handling methods were followed for each assay, the mean viral load results from PPTs evaluated in the studies only differed by a range of -0.03 to +0.08 log copies/ml from the mean viral load results from standard EDTA plasma. This difference is well-within the standard variation seen in viral load assays even when the same sample is tested twice, and far below the 0.3–0.5 log copies/ml difference that is considered clinically significant [29,30].

Tab. 3. Published handling methods for commercially available PPTs and viral load assays.
Published handling methods for commercially available PPTs and viral load assays.

Discussion

PPTs allow plasma to be prepared, stored, and transported in the same tube used to collect venous whole blood. PPTs have been shown to provide equivalent viral load results to plasma from standard EDTA tubes if proper handling of PPTs is followed according to the guidance from independently published studies included in this review as well as manufacturer instructions [31,32].

Regardless of assay type used, centrifugation of PPTs and/or aliquoting of plasma into a secondary tube prior to viral load testing has also been consistently shown to prevent falsely elevated viral load results. Performing these steps will ensure that the cellular component of whole blood has been completely separated from the plasma in order to prevent HIV proviral DNA, intracellular HIV RNA, and HIV particles associated with platelets from falsely elevating the viral load results [33,34]. Assays that are more HIV RNA specific are likely to be less impacted by the presence of proviral DNA and may explain why some assays still show accurate viral load results even when PPTs are frozen and thawed in situ, without a separate aliquoting or centrifugation step.

PPTs can improve access to viral load testing in several ways. The ability to store plasma for longer periods of time than in standard EDTA tubes, as indicated by an earlier systematic review [9], could expand the reach of viral load laboratories, which is particularly useful in resource limited settings. Additionally, easier sample handling with fewer manual steps and a reduced risk of sample contamination if the plasma is kept in the PPT can streamline laboratory workflows and increase testing throughput. Although they are typically more expensive than EDTA tubes, if PPTs are available and staff can be trained on their use, they provide another option for HIV programs deciding how to expand coverage of viral load testing. Laboratories using PPTs must then be familiar with proper handling methods and any additional equipment (e.g. secondary tubes for aliquoting, centrifugation needs prior to testing) needed to ensure accurate results. Existing procurement and supply chain mechanisms which have simplified purchasing of phlebotomy supplies for many countries through all-in-one sample bundles would also need to be updated to include PPTs, since they typically only include EDTA tubes currently. Table 4 summarizes the advantages and challenges associated with using PPTs for viral load testing identified through this review.

Tab. 4. Advantages and challenges associated with PPTs.
Advantages and challenges associated with PPTs.

A major limitation of this review is that it does not encompass all PPTs and viral load tests that are currently commercially available. All of the studies in this review evaluated only BD Vacutainer® PPTTMs, the most commonly used PPT brand. Other PPTs, also referred to as EDTA with gel separator tubes, are commercially available from manufacturers such as Grenier and TUD [35,36]. Additionally, other viral load assays such as the Cepheid Xpert HIV-1 and Hologic Aptima HIV-1 Quant Dx assays are the only other tests, in addition to the Abbott and Roche viral load assays, that are currently commercially available with regulatory approval for the use of PPTs; however, their instructions for use do not provide any further specific guidance on how PPTs should be used [37,38]. No independent studies were found evaluating other brands of PPTs or the use of PPTs with these viral load assays.

While the majority of studies were done in North America or Western Europe, the studies still examined a wide range of storage and transport conditions that could be found anywhere in the world indicating generalizability of the results. Additionally, when PPTs were handled properly, the results were accurate regardless of the population in which the study was conducted. Scientifically, PPTs would not be expected to separate plasma differently in viral load laboratories in high burden HIV countries, and their gel separation capability would not be expected to be affected by any known patient characteristics. However, it is important to note that methods and storage/shipping times were simulated in the laboratory, rather than studied in decentralized settings. Additionally, further research on other brands of PPTs as well as with other viral load assays that have regulatory approval would be helpful to confirm the findings of the studies in this review.

Overall, this review summarized results from 16 studies of PPTs involving results from over 6,000 individuals. Consideration of PPTs may be particularly worthwhile in settings where simpler plasma preparation, reduced cross-contamination risk, and longer sample transport times can facilitate the scale up of viral load testing. PPTs will add to the menu of options available to countries to expand access to viral load, particularly in settings that may require more support and extended shipping or storage times compared to standard EDTA samples. Increasing access to viral load will improve patient care and services, while also supporting country goals to reach the 90-90-90 targets.

Supporting information

S1 File [docx]
Search terms protocol.

S1 Table [docx]
PRISMA checklist.

S2 Table [docx]
QUADAS-2 assessment of individual studies.


Zdroje

1. World Health Organization. Consolidated guidelines on the use of antiretroviral drugs for treating and preventing HIV infection: recommendations for a public health approach. 2nd ed. Geneva: World Health Organization. 2016.

2. UNAIDS. 90-90-90: An ambitious treatment target to help end the AIDS epidemic. Geneva: UNAIDS. 2014.

3. World Health Organization. Technical and operational considerations for implementing HIV viral load testing. Geneva: World Health Organization. 2014.

4. Médecins Sans Frontières. Putting HIV and HCV to the test: a product guide for point-of-care CD4 and laboratory-based and point-of-care HIV and HCV viral load tests. Geneva: MSF. 2017.

5. Habiyambere V, Dongmo Nguimfack B, Vojnov L, Ford N, Stover J, Hasek L, et al. Forecasting the global demand for HIV monitoring and diagnostic tests: A 2016–2021 analysis. PLoS One. 2018 Sep 19;13(9):e0201341. doi: 10.1371/journal.pone.0201341 30231022

6. Clinton Health Access Initiative. HIV market report. Boston: CHAI. 2018. Available from https://clintonhealthaccess.org/content/uploads/2018/09/2018-HIV-Market-Report_FINAL.pdf

7. Smit PW, Sollis KA, Fiscus S, Ford N, Vitoria M, Essajee S, et al. Systematic review of the use of dried blood spots for monitoring HIV viral load and for early infant diagnosis. PLoS One. 2014 Mar 6;9(3):e86461. doi: 10.1371/journal.pone.0086461 24603442

8. Becton, Dickinson, and Company. BD Vacutainer® PPT Instructions for Use. 2015. Available from https://www.bdbiosciences.com/ds/ab/others/PI_PPT_March_2016_VDP40162-03_Web.pdf

9. Bonner K, Siemieniuk RA, Boozary A, Roberts T, Fajardo E, Cohn J. Expanding access to HIV viral load testing: A systematic review of RNA stability in EDTA tubes and PPT beyond current time and temperature thresholds. PLoS One. 2014 Dec 1;9(12):e113813. doi: 10.1371/journal.pone.0113813 25437009

10. Moher D, Liberati A, Teztlaff J, Altman DG, PRISMA Group. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. PloS Med. 2009;6:e1000097. doi: 10.1371/journal.pmed.1000097 19621072

11. Whiting PF, Rutjes AW, Westwood ME, Mallett S, Deeks JJ, Reitsma JB, et al. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med. 2011;155:529–536. doi: 10.7326/0003-4819-155-8-201110180-00009 22007046

12. Schünemann HJ, Oxman AD, Brozek J, Glasziou P, Jaeschke R, Vist GE, et al. Grading quality of evidence and strength of recommendations for diagnostic tests and strategies. BMJ. 2008;336:1106–1110. doi: 10.1136/bmj.39500.677199.AE 18483053

13. Holodniy M, Mole L, Yen-Lieberman B, Margolis D, Starkey C, Carroll R, et al. Comparative stabilities of quantitative human immunodeficiency virus RNA in plasma from samples collected in VACUTAINER CPT, VACUTAINER PPT, and standard VACUTAINER tubes. J Clin Microbiol. 1995 Jun;33(6):1562–6. 7650187

14. Ginocchio CC, Wang XP, Kaplan MH, Mulligan G, Witt D, Romano JW, et al. Effects of specimen collection, processing, and storage conditions on stability of human immunodeficiency virus type 1 RNA levels in plasma. J Clin Microbiol. 1997 Nov;35(11):2886–93. 9350753

15. Holodniy M, Rainen L, Herman S, Yen-Lieberman B. Stability of plasma human immunodeficiency virus load in VACUTAINER PPT plasma preparation tubes during overnight shipment. J Clin Microbiol. 2000 Jan;38(1):323–6. 10618109

16. Elbeik T, Nassos P, Kipnis P, Haller B, Ng VL. Evaluation of the VACUTAINER PPT Plasma Preparation Tube for use with the Bayer VERSANT assay for quantification of human immunodeficiency virus type 1 RNA. J Clin Microbiol. 2005 Aug;43(8):3769–71. doi: 10.1128/JCM.43.8.3769-3771.2005 16081908

17. Giordano M, Kelleher T, Colonno RJ, Lazzarin A, Squires K. The effects of the Roche AMPLICOR HIV-1 MONITOR UltraSensitive Test versions 1.0 and 1.5 viral load assays and plasma collection tube type on determination of response to antiretroviral therapy and the inappropriateness of cross-study comparisons. J Clin Virol. 2006 Apr;35(4):420–5. 16604577

18. Griffith BP, Mayo DR. Increased levels of HIV RNA detected in samples with viral loads close to the detection limit collected in Plasma Preparation Tubes (PPT). J Clin Virol. 2006 Feb;35(2):197–200. 16253548

19. Wan H, Seth A, Rainen L, Fernandes H. Coamplification of HIV-1 proviral DNA and viral RNA in assays used for quantification of HIV-1 RNA. J Clin Microbiol. 2010 Jun;48(6):2186–90. doi: 10.1128/JCM.02034-09 20181913

20. Salimnia H, Moore EC, Crane LR, Macarthur RD, Fairfax MR. Discordance between viral loads determined by Roche COBAS AMPLICOR human immunodeficiency virus type 1 monitor (version 1.5) Standard and ultrasensitive assays caused by freezing patient plasma in centrifuged becton-dickinson vacutainer brand plasma preparation tubes. J Clin Microbiol. 2005 Sep;43(9):4635–9. doi: 10.1128/JCM.43.9.4635-4639.2005 16145119

21. García-Bujalance S, Ladrón de Guevara C, González-García J, Arribas JR, Zamora F, Gutiérrez A. Elevation of viral load by PCR and use of plasma preparation tubes for quantification of human immunodeficiency virus type 1. J Microbiol Methods. 2007 May;69(2):384–6. doi: 10.1016/j.mimet.2007.01.018 17363096

22. Rebeiro PF, Kheshti A, Bebawy SS, Stinnette SE, Erdem H, Tang YW, et al. Increased detectability of plasma HIV-1 RNA after introduction of a new assay and altered specimen-processing procedures. Clin Infect Dis. 2008 Nov 15;47(10):1354–7. doi: 10.1086/592693 18922071

23. Kran AM, Jonassen TØ, Sannes M, Jakobsen K, Lind A, Maeland A, et al. Overestimation of human immunodeficiency virus type 1 load caused by the presence of cells in plasma from plasma preparation tubes. J Clin Microbiol. 2009 Jul;47(7):2170–4. doi: 10.1128/JCM.00519-09 19420166

24. Kraft CS, Binongo JN, Burd EM, Eaton ME, McCloskey CB, Fernandes H, et al. Successful use of Plasma Preparation Tubes™ (PPTs) in the COBAS® AmpliPrep/COBAS® TaqMan® HIV-1 test. J Clin Virol. 2013 May;57(1):77–9. doi: 10.1016/j.jcv.2012.12.015 23332979

25. Fernandes H, Morosyuk S, Abravaya K, Ramanathan M, Rainen L. Evaluation of effect of specimen-handling parameters for plasma preparation tubes on viral load measurements obtained by using the Abbott RealTime HIV-1 load assay. J Clin Microbiol. 2010 Jul;48(7):2464–8. doi: 10.1128/JCM.00083-10 20484602

26. Adachi D, Benedet M, Tang JW, Taylor GD. Comparative evaluation of Roche's COBAS Ampliprep/COBAS TaqMan HIV-1 Test v2.0 and Abbott's RealTime m2000sp/rt HIV-1 assay on PPTs and EDTA samples. J Clin Virol. 2014 May;60(1):78–9. doi: 10.1016/j.jcv.2014.01.019 24594081

27. Cloherty G, Swanson P, Lucic D, Dieckhaus K, Anthony P, Cataline P, et al. Clinical implications of elevated HIV-1 viral load results obtained from samples stored frozen in vacutainer plasma preparation tubes. J Virol Methods. 2014 Aug;204:91–2. doi: 10.1016/j.jviromet.2014.01.025 24747107

28. Goedhals D, Scott LE, Moretti S, Cooper MA, Opperman WJ, Rossouw I. Evaluation of the use of plasma preparation tubes for HIV viral load testing on the COBAS AmpliPrep/COBAS TaqMan HIV-1 version 2.0. J Virol Methods. 2013 Feb;187(2):248–50. doi: 10.1016/j.jviromet.2012.11.019 23178587

29. Sollis KA, Smit PW, Fiscus S, Ford N, Vitoria M, Essajee S, et al. Systematic review of the performance of HIV viral load technologies on plasma samples. PLoS One. 2014 Feb 18;9(2):e85869. doi: 10.1371/journal.pone.0085869 24558359

30. Modjarrad K, Chamot E, Vermund SH. Impact of small reductions in plasma HIV RNA levels on the risk of heterosexual transmission and disease progression. AIDS. 2008 Oct 18;22(16):2179–85. doi: 10.1097/QAD.0b013e328312c756 18832881

31. Becton, Dickinson, and Company. BD White Paper VS8188: Evaluation Of The Effect Of Specimen Handling Conditions In BD Vacutainer® PPT On The Stability Of HIV-1 Viral Load Using Roche Cobas® Ampliprep/Cobas® Taqman® HIV-1 Test. 2010. Available from https://cbpwhitepapers.bd.com/bdvacutainerppttube.aspx

32. Becton, Dickinson, and Company. BD White Paper VS8189: Evaluation of Specimen Handling Conditions in BD Vacutainer® Plasma Preparation Tube HIV-1 Viral Load as Measured by the Abbott RealTime HIV-1 Assay. 2010. Available from https://cbpwhitepapers.bd.com/bdvacutainerppttube.aspx

33. Sabino EC, Gaburo N Jr, Leite JP, Viecili M, Rosenfeld LG. Platelets may affect detection and quantitation of HIV RNA in plasma samples with low viral loads. J Acquir Immune Defic Syndr. 2004 Nov 1;37(3):1432–3. doi: 10.1097/00126334-200411010-00013 15483474

34. Lee TH, Stromberg RR, Heitman JW, Sawyer L, Hanson CV, Busch MP. Distribution of HIV type 1 (HIV-1) in blood components: detection and significance of high levels of HIV-1 associated with platelets. Transfusion. 1998 Jun;38(6):580–8. doi: 10.1046/j.1537-2995.1998.38698326338.x 9661692

35. Greiner Bio-One. Evaluation of VACUETTE® K2EDTA Gel Tubes for Molecular Diagnostics. 2009. Available from https://www.gbo.com/fileadmin/user_upload/Downloads/White_Papers/WP_Evaluation_K2Gel_USA09_Rev00.pdf

36. TUD. Products Description: K2 EDTA with Gel Tubes. 2010. Available from http://www.tud.my/file/TUD_Leaflet_K2EDTA%20Gel_White_R1_FAol.pdf

37. WHO Prequalification of Diagnostics Programme. Public Report: Xpert® HIV-1 Viral Load. 2017. Available at https://www.who.int/diagnostics_laboratory/evaluations/pq-list/hiv-vrl/170830_amended_final_pqpr_0192_0193_0194_0195_070_00_v3.pdf?ua=1

38. WHO Prequalification of Diagnostics Programme. Public Report: AptimaTM HIV-1 Quant Dx Assay. 2018. Available at https://www.who.int/diagnostics_laboratory/evaluations/pq-list/180207_final_pqpr_0236_078_00_v5_IFU_v4.pdf?ua=1


Článok vyšiel v časopise

PLOS One


2019 Číslo 11
Najčítanejšie tento týždeň
Najčítanejšie v tomto čísle
Kurzy

Zvýšte si kvalifikáciu online z pohodlia domova

Aktuální možnosti diagnostiky a léčby litiáz
nový kurz
Autori: MUDr. Tomáš Ürge, PhD.

Všetky kurzy
Prihlásenie
Zabudnuté heslo

Zadajte e-mailovú adresu, s ktorou ste vytvárali účet. Budú Vám na ňu zasielané informácie k nastaveniu nového hesla.

Prihlásenie

Nemáte účet?  Registrujte sa

#ADS_BOTTOM_SCRIPTS#