A psychoacoustic application for the adjustment of electrical hearing thresholds in cochlear implant patients
Autoři:
Johannes Plesch aff001; Benjamin P. Ernst aff002; Sebastian Strieth aff002; Tobias Rader aff001
Působiště autorů:
Division of Audiological Acoustics, Department of Otolaryngology, University Medical Center, Mainz, Germany
aff001; Department of Otolaryngology—Head and Neck Surgery, University Medical Center, Mainz, Germany
aff002
Vyšlo v časopise:
PLoS ONE 14(10)
Kategorie:
Research Article
prolekare.web.journal.doi_sk:
https://doi.org/10.1371/journal.pone.0223625
Souhrn
Objective
Fitting cochlear implants, especially the precise determination of electrical hearing thresholds, is a time-consuming and complex task for patients as well as audiologists. Aim of the study was to develop a method that enables cochlear implant (CI) patients to determine their electrical hearing thresholds precisely and independently. Applicability and impact of this method on speech perception in noise at soft speech levels were evaluated.
Method
An adaptive psychoacoustic procedure for precise hearing threshold determination (precT) was implemented using MatLab (MathWorks) and a graphical user interface was created. Sound signals were calibrated with a CIC4-Implant-Decoder. Study design: A prospective study including 15 experienced adult cochlear implant users was conducted. Electrical hearing thresholds were determined using the automated precT procedure (auto-precT). Speech perception in noise at 50 dB SPL presentation levels was measured for three conditions: (P1) T-levels kept at the previously established T-levels; (P2) T-levels set to the hearing thresholds determined using auto-precT application; (P3) T-levels set 10 cu below the values determined with auto-precT.
Results
All subjects were able to perform the auto-precT application independently. T-levels were altered on average by an absolute value of 10.5 cu using auto-precT. Median speech reception thresholds were significantly improved from 2.5 dB SNR (P1) to 1.6 dB SNR (P2, p = 0.02). Speech perception was lowest using the globally lowered T-levels, median 2.9 dB SNR (P3, not significant compared to P1 and P2).
Conclusion
The applicability of the developed auto-precT application was confirmed in the present clinical study. Patients benefited from adjusting previously established T-levels to the threshold levels determined by the auto-precT application. The integration of the application in the clinical fitting routine as well as a remote fitting software approach is recommended. Furthermore, future possibilities of auto-precT include the implementation of the application on tablets or smart phones.
Klíčová slova:
Diet – Medical implants – Sensory perception – Functional electrical stimulation – Audio signal processing – Audiology – Computer software – Hearing
Zdroje
1. Vaerenberg B, Smits C, De Ceulaer G, Zir E, Harman S, Jaspers N, et al. Cochlear implant programming: a global survey on the state of the art. ScientificWorldJournal. 2014;2014:501738. doi: 10.1155/2014/501738 24688394; PubMed Central PMCID: PMC3932199.
2. Firszt JB, Holden LK, Skinner MW, Tobey EA, Peterson A, Gaggl W, et al. Recognition of Speech Presented at Soft to Loud Levels by Adult Cochlear Implant Recipients of Three Cochlear Implant Systems. Ear & Hearing. 2004;25(4):375–87.
3. Holden LK, Reeder RM, Firszt JB, Finley CC. Optimizing the perception of soft speech and speech in noise with the Advanced Bionics cochlear implant system. Int J Audiol. 2011;50(4):255–69. doi: 10.3109/14992027.2010.533200 21275500; PubMed Central PMCID: PMC3434686.
4. Skinner MW, Holden LK, Holden TA, Demorest ME. Comparison of two methods for Selecting Minimum Stimulation Levels Used in Programming the Nucleus 22 Cochlear Implant. Journal of Speech and Hearing Research. 1999;42(4):814–28.
5. Willeboer C, Smoorenburg GF. Comparing Cochlear Implant Users’ Speech Performance with Processor Fittings Based on Conventionally Determined T and C Levels or on Compound Action Potential Thresholds and Live-Voice Speech in a Prospective Balanced Crossover Study. Ear & Hearing. 2006;27(6):789–98.
6. Botros A, Banna R, Maruthurkkara S. The next generation of Nucleus((R)) fitting: a multiplatform approach towards universal cochlear implant management. Int J Audiol. 2013;52(7):485–94. doi: 10.3109/14992027.2013.781277 23617610; PubMed Central PMCID: PMC3696341.
7. Mewes A, Hey M. Einfluss der T-Level auf das Sprachverstehen in Ruhe und im Störschall bei erwachsenen CI-Patienten. Conference Paper, 20 Jahrestagung der Deutschen Gesellschaft für Audiologie. 2017.
8. Skinner MW, Holden LK, Holden TA, Demorest ME. Comparison of Procedures for Obtaining Thresholds and Maximum Acceptable Loudness Levels With the Nucleus Cochlear Implant System. Journal of Speech and Hearing Research. 1995;38:677–89. doi: 10.1044/jshr.3803.677 7674659
9. van Wieringen A, Wouters J. Comparison of Procedures to Determine Electrical Stimulation Thresholds in Cochlear Implant Users. Ear & Hearing. 2001;22(No. 6):528–38
10. Plant K, Law M, Whitford L, Knight M, Tari S, Leigh J, et al. Evaluation of Streamlined Programming Procedures for the Nucleus Cochlear Implant with the Contour Electrode Array. Ear & Hearing. 2005;26(6):651–68.
11. Rader T, Doms P, Adel Y, Weissgerber T, Strieth S, Baumann U. A method for determining precise electrical hearing thresholds in cochlear implant users. Int J Audiol. 2018:1–8. Epub 2018/02/03. doi: 10.1080/14992027.2017.1412519 29390897.
12. Skinner MW, Holden LK, Whitford LA, Plant KL, Psarros C, Holden TA. Speech Recognition with the Nucleus 24 SPEAK, ACE, and CIS Speech Coding Strategies in Newly Implanted Adults. Ear & Hearing. 2002b;23(3):207–223. doi: 10.1097/00003446-200206000-00005 12072613
13. Bewley MS. Mining clinical databases: A post-hoc study of cochelar implant fitting practices. Cochlear White Paper. 2013.
14. Arndt P, Staller S, Arcaroli J, Hines A, Ebinger K. Within-subject comparison of advanced coding strategies in the Nucleus 24 cochlear implant. Cochlear White Paper. 1999.
15. Busby PA, Arora K. Effects of Threshold Adjustment on Speech Perception in Nucleus Cochlear Implant Recipients. Ear Hear. 2016;37(3):303–11. Epub 2015/12/17. doi: 10.1097/AUD.0000000000000248 26671316.
Článok vyšiel v časopise
PLOS One
2019 Číslo 10
- Metamizol jako analgetikum první volby: kdy, pro koho, jak a proč?
- Nejasný stín na plicích – kazuistika
- Masturbační chování žen v ČR − dotazníková studie
- Je Fuchsova endotelová dystrofie rohovky neurodegenerativní onemocnění?
- Fixní kombinace paracetamol/kodein nabízí synergické analgetické účinky
Najčítanejšie v tomto čísle
- Correction: Low dose naltrexone: Effects on medication in rheumatoid and seropositive arthritis. A nationwide register-based controlled quasi-experimental before-after study
- Combining CDK4/6 inhibitors ribociclib and palbociclib with cytotoxic agents does not enhance cytotoxicity
- Prevalence of pectus excavatum (PE), pectus carinatum (PC), tracheal hypoplasia, thoracic spine deformities and lateral heart displacement in thoracic radiographs of screw-tailed brachycephalic dogs
- Risk factors associated with IgA vasculitis with nephritis (Henoch–Schönlein purpura nephritis) progressing to unfavorable outcomes: A meta-analysis